All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Fully Autoclave:
The MicroPette plus pipettes can be fully autoclaved at 121ºC, 1 atm for minutes. The pipettes can be autoclaved without special preparations. After autoclaving the pipette must be cooled down and dried over 12 hours.
Features for MicroPette plus:
* Fully autoclavable
* Light weight, ergonomic, low force design
* Digital display clearly reads volume setting
* The pipettes cover volume range of 0.1μl to 10ml
* Easy to calibrate and maintain
* Design helps avoid repetitive strain injuries
* Manufactured from innovative materials
Features for Multi-channel pipette
* 8 and 12 channel pipettes are available for standard 96-well plate
* Individual piston and tip cone assemblies allow easy repair and maintenance
* Compound material tip cone design allows visual seal vertification
* Universal pipette tips can be used
* Dispensing head rotates ensure optimum pipetting convenience
Single-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume (μl) | Maximume Permissible Systematic error (Inaccuracy) | Maximume Permissible Random error (Inaccuracy) | ||
% | μl | % | μl | |||
0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | ||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | ||
0.5-10μl | 0.1μl | 10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | ||
1 | 2.50% | 0.025 | 1.50% | 0.015 | ||
2-20μl | 0.5μl | 20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | ||
2 | 3.00% | 0.06 | 2.00% | 0.04 | ||
5-50μl | 0.5μl | 50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | ||
5 | 2.00% | 0.1 | 2.00% | 0.1 | ||
10-100μl | 1μl | 100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
10 | 3.00% | 0.3 | 1.50% | 0.15 | ||
20-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
20 | 3.00% | 0.6 | 1.00% | 0.2 | ||
50-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | ||
50 | 1.00% | 0.5 | 0.40% | 0.2 | ||
100-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
100 | 2.00% | 2 | 0.70% | 0.7 | ||
200-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | ||
200 | 0.90% | 1.8 | 0.30% | 0.6 | ||
1000-5000μl | 50μl | 5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | ||
1000 | 0.70% | 7 | 0.30% | 3 |
8-channel Adjustable Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) | Maximume Permissible Systematic error (Inaccuracy) | Maximume Permissible Random error (Inaccuracy) | ||||
% | μl | % | μl | |||||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 | ||
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 | ||
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 | ||
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 | ||||
12-channel Adjustable Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) | Maximume Permissible Systematic error (Inaccuracy) | Maximume Permissible Random error (Inaccuracy) | ||||
% | μl | % | μl | |||||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 | ||
5 | 2.50% | 0.125 | 2.50% | 0.125 | ||||
1 | 4.00% | 0.04 | 4.00% | 0.04 | ||||
5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 | ||
25 | 1.50% | 0.375 | 1.00% | 0.25 | ||||
5 | 3.00% | 0.15 | 2.00% | 0.1 | ||||
50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 | ||
150 | 1.00% | 1.5 | 0.50% | 0.75 | ||||
50 | 1.50% | 0.75 | 0.80% | 0.4 | ||||
Fixed Volume Pipettes | ||||||||
Volume Range | Increment | Test Volume (μl) | Maximume Permissible Systematic error (Inaccuracy) | Maximume Permissible Random error (Inaccuracy) | ||||
% | μl | % | μl | |||||
5μl | - | 5μl | 1.3% | 0.065 | 1.2% | 0.06 | ||
10μl | - | 10μl | 0.8% | 0.08 | 0.8% | 0.08 | ||
20μl | - | 20μl | 0.6% | 0.12 | 0.5% | 0.1 | ||
25μl | - | 25μl | 0.5% | 0.125 | 0.3% | 0.075 | ||
50μl | - | 50μl | 0.5% | 0.25 | 0.3% | 0.15 | ||
100μl | - | 100μl | 0.5% | 0.5 | 0.3% | 0.3 | ||
200μl | - | 200μl | 0.4% | 0.8 | 0.2% | 0.4 | ||
250μl | -- | 250μl | 0.4% | 1.0 | 0.2% | 0.5 | ||
500μl | - | 500μl | 0.3% | 1.5 | 0.2% | 1.0 | ||
1000μl | - | 1000μl | 0.3% | 3.0 | 0.2% | 2.0 | ||
2000μl | - | 2000μl | 0.3% | 6.0 | 0.15% | 3.0 | ||
5000μl | - | 5000μl | 0.3% | 15 | 0.15% | 7.5 |
Water activity meter, indicates the free water content when food is stable, measures the possibility of microbial proliferation.
Food formulators use water activity to predict how much moisture migration will affect their product, so that to choose suitable package and storage method to reduce the usage of antiseptics.