1.Conductivity measuring range: 0.01~20μS/cm (Electrode: K=0.01) |
0.1~200μS/cm (Electrode: K=0.1) |
1.0~2000μS/cm (Electrode: K=1.0) |
10~20000μS/cm (Electrode: K=10.0) |
30~600.0mS/cm (Electrode: K=30.0) |
2. Intrinsic error of the electronic unit: conductivity: ±0.5%FS, temperature: ±0.30ºC |
3. Range of automatic temperature compensation: 0~99.9ºC, with 25ºC as the reference temperature;0~150ºC (optional) |
4. Water sample tested: 0~99.9ºC, 0.6MPa |
5. Intrinsic error of the instrument: conductivity: ±1.0%FS, temperature: ±0.5ºC |
6. Automatic temperature compensation error of the electronic unit: ±0.5%FS |
7. Repeatability error of the electronic unit: ±0.2%FS±1 Unit |
8. Stability of the electronic unit: ±0.2%FS±1 unit/24h |
9. Isolated current output: 0~10mA ( load<1.5kΩ) |
4~20mA (load<750Ω) Can double the current output. |
10. Output current error: ≤±l%FS |
11. Error of electronic unit caused by ambient temperature: ≤±0.5%FS |
12. Error of the electronic unit caused by the supply voltage: ≤±0.3%FS |
13. Alarm relay: AC 220V, 3A |
14. Communication interface: RS485 (optional) |
15. Power supply: AC 220V±22V, 50Hz±1Hz, DC 24V(optional) |
16. Protection grade: IP65 |
17. Clock accuracy: ±1 minute/month |
18. Data storage capacity: l month (1 point/5 minutes) |
19. Saving time of data under a continuous power-failure condition: 10 years |
20. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm; dimension of the hole: 138 x 138mm |
21. Working conditions: ambient temperature: 0~60ºC; relative humidity <85% |
22. Weight: 0.6kg |
23. The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, and 30.0. |