OEM Factory Price and Quality Hospital Instrument of Oxygen N2o Nitrous Oxide Gas Reduce Box

Min.Order: 10
Product origin: Ningbo, Zhejiang, China
Infringement complaint: complaintComplaint
US$ 700 ~ 1000

Description

Product Description

Second grade pressure reducing valve box is also integrated with the pressure pipe and cut-off function, the purpose of decompression is pipeline pressure control in a certain pressure range, such as for ward the oxygen pressure control in about 0.3~0.5Mpa for hyperbaric oxygen chamber pressure control in about 0.6~ 0.8MPa and from the bus or liquid oxygen tank output pressure is generally in the 0.8~1.0Mpa left and right.
In the use of gas in the region to set up two level decompression device, but also can play the role of buffer pipe pressure, reduce the impact on the patient, reduce the application risk. Vacuum suction system and AGSS system without the need for decompression, and oxygen, compressed air, nitrous oxide and other medical gases generally need to set the second grade pressure reducing device in the pipeline. The function of the main function is to achieve the opening and closing of the pipeline.

  1. Input pressure:0.8-1.2MPa
  2. Output pressure:0.35-0.5MPa
  3. Medium:O2
  4. Flow rate:400L/min
  5. Safety pressure:0.52MPa
  6. Size:45*18*55cm
  7. With alarm or without
  8. With flowmeter or without