Anesthesia Machine | |||||
Model | THR-MJ-560B1 | THR-MJ-560B2 | THR-MJ-560B3 | THR-MJ-560B4 | THR-MJ-560B5 |
Part of Anesthesia Machine | |||||
Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistan | ||||
Scope of application | adult, child(4 years above) | ||||
Gas source | O2:0.28~0.6 MPa | O2, N2O: 0.28~0.6MPa | O2, N2O, Air: 0.28~0.6MPa | ||
Flow meter | O2: 0.05-1.0L/min 1.1-10L/min | O2,N2O: 0.05- 10L/min 0.05-10L/min | O2,N2O: 0.05-1.0L/min 1 .1-10L/min | O2,N2O: 0.05-1.0L/min 1.1-10L/min | O2,N2O: 0.05-1.0L/min 1.1-10L/min Air:0. 05-10L/min |
O2, N2O linkage and N2O Stopper | / | When using nitrous oxide, oxygen concentration>25%. When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. | |||
Vaporizer | One, high-precision evaporator | One, high-precision evaporator | One, high-precision evaporator | Two, high-precision evaporator | Two,high-precision evaporator |
Part of Ventilator | |||||
Display mode | 5.7" | 5.7" | 10.4" TFT | 5.7" | 10.4" TFT |
mode | IPPV, SIPPV, SIMV, MANUAL | IPPV,SIPPV, SIMV, MANUAL | IPPV, SIPPV, SIMV, MANUAL, VCV, PCV | IPPV,SIPPV,SIMV, MANUAL | IPPV, SIPPV, SIMV, MANUAL, VCV, PCV |
function | PEEP, SIGH, IRV | PEEP, SIGH, IRV | PEEP,SIGH, IRV, IP, standby | PEEP, SIGH, IRV | PEEP,SIGH, IRV, IP, standby |
TV | 50~1500ml | 50~1500ml | 20~1600 ml | 50~1500ml | 20~1600 ml |
Rate | 2~99 bpm | 2~99 bpm | 2~99 bpm | 2~99 bpm | 2~99 bpm |
SIMV rate | 2~20 bpm | 2~20 bpm | 2~20 bpm | 2~20 bpm | 2~20 bpm |
I:E | 2:1~1:8 | 2:1~1:8 | 4:1~1:8 | 2:1~1:8 | 4:1~1:8 |
Inspiratory trigger pressure | -10~20cmH2O | -10~20cmH2O | -10~20cmH2O | -10~20cmH2O |
-10~20cmH2O |
PEEP | 2~20cmH2O | 2~20cmH2O | 2~20cmH2O | 2~20cmH2O | 2~20cmH2O |
Pressure limitation | 10~60cmH2O | 5~60 cmH2O | 10~60cmH2O | 5~60 cmH2O | |
SIGH | 1.5 times the inspiratory time/100bpm | SIGH 1.5 times the inspiratory time (60-120 adjustable) | 1.5 times the inspiratory time/100bpm | IGH 1.5 times the inspiratory time (60-120 adjustable) | |
Inspiratory platform | / | / | 0~50% | / | 0~50% |
Pressure control | / | / | 3~50 cmH2O | / | 3~50 cmH2O |
Pressure slope | / | / | 1~10 gear | / | 1~10 gear |
Power | AC 220V 50Hz (UPS, Storage battery) | ||||
Parameters for ventilation monitoring |
B1, B2, B4: tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow - time waveform, PEEP, inspiratory trigger pressure
B3, B5: tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, inhale tidal volume, inspiratory platform, average pressure, autonomous inspiratory rate, oxygen concentration, battery capacity, flow-volume loop, pressure-volume loop | ||||
Mechanical Arms | / | / | YES | YES | YES |
End-tidal CO2 monitoring | Optional |