Description:
It has high performance neutral electrode contact inspection system, which can detect electrode contact area, and system will test and evaluate the effective contact area between electro and derma, if the contact area was found low to a danger level, system will automatic cutout and gives alarm indication.
High definition large LED indication, with audio & visual indication.
Main unit has automatic monitoring function and mistake indication function.
Seven output mode:
Output Rating:
a) Cut: 400W(Load 500Ω)
b)Blend 1: 200W(Load 500Ω)
c)Blend 2: 150W(Load 500Ω)
d)Electro Coag 1: 120W(Load 500Ω)
e)Electro Coag 2: 80W (Load 500Ω)
f)Bipolar Coag 1: 50W (Load 200Ω)
g) Argon: Dual Argon airflow adjustment, accuracy: 0.1 L/min;
Power: 220V±22V, 50Hz±1Hz
Technical Parameters:
1. Ambient temp. range: 5ºC~40 ºC 2. Relative Humidity Range: ≤80
3. Atmospheric Pressure Range: 86.0~106.0kpa 4. Power: 110-220V±22V, 50-60Hz±1
5. Operating Freq.: 330~460kHz 6. Rated Power:880VA±10%
7. Type: CF
Power Consumption: ≤1100VA(Cut function 400W)
Transportation and Storage:
a) Ambient Temp. Range: -40 ºC~55 ºC b) Relative Humidity: ≤95%
c) Atmospheric Pressure: 500hpa~1060hpa
Application:
Applicable for general surgery, thoracic surgery, orthopedics, obstetrics and gynecology, urologic surgery, cerebral surgery, microsurgery, ENT surgery and etc.
Cut: Pure cut, can make easy operation for any organic, the effect is refine cut,
no escharosis on the surface.
Blend 1: The effect is cut with coagulation, Can use this function together with Endoscope.
Blend 2: The effect is offering a better coagulation during cut, Can use this function together with Argon.
Electro Coag 1: Called strength electrocoagulation, or called surface coagulation, for a deep coagulation layer for organic, small effect area, Can use this function together with Endoscope..
Electro Coag2: Called spray electrocoagulation, offers effective non-contact coagulation, Can
use this function together with Argon.
Bipolar Coag 1: Called rapid Bipolar coagulation, to provide coagulation function rapidly at boot,which can reduce mis-operation rate during operation.
Argon: Dual Argon airflow adjustament, accuracy: 0.1 L/min;
Argon pressure range: 0.20 ~ 0.50 Mpa.
Can cut underwater, suitable for use with Argon system, Endoscope, uteroscope etc.
including organic cut in rich fat area.
Standard Configuration
Main unit: 1 pc
Double foot switch: 1 pc
ESU pencil: 2 pcs
Electrode (Disposable): 5 pcs
Bipolar forceps: 1 pc
Bipolar forceps cable: 1 pc
Electrode cable: 1 pc
Power cable: 1 pc
Fuse: 2 pcs
User's manual: 1 pc
Packing List: 1 pc
Guarantee card: 1 pc
.