Item | Content | Technical Parameters | |
Power | Voltage | 380V±38V | |
Frequency | 50Hz±1Hz | ||
Capacity | ≥63kVA | ||
Internal Resistance | ≤0.17Ω | ||
High Frequency High Voltage X-Ray Generator (SEDECAL) | Power | 50KW | |
Inverter Frequency | 60 KHz | ||
Tube voltage | 40kV-150kV ,continuous adjustment | ||
Tube current | 10mA-640mA , grade adjustment | ||
Exposure time | 0.1ms-10000ms | ||
X-ray source assembly (TOSHIBA E7884X) | Tube Focus:big/small | 1.2mm /0.6mm | |
Input power | Big focus 50kW Small focus 20kW | ||
Anode thermal capacity | 210KJ | ||
Rotary anode speed | 2700rpm | ||
You can choose one kind of Flat Panel Detector from Toshiba and Vieworks, the information as belows: | |||
Toshiba Digital Flat Panel Detector | Toshiba | FDX4343R | |
Active area | 430(H)×439(V) | ||
Pixel matrix: | 3008(H)×3072(V) | ||
Pixel pitch: | 143 μm | ||
Cycle time: | Less than 6 s | ||
Limiting resolution | 3.5 lp/mm typ | ||
A / D transition: | 14 bit | ||
Energy range | 40 - 150 kVp | ||
Maximum entrance dose (low gain) | 4 mR / frame | ||
Data output | 16-bit digital output Ethernet(1000BASE - T) | ||
Exposure Synchronous control: | outside | ||
power Input: | DC 24V 2A | ||
Digital Flat Penal Detector | Active area | 350(H)×430(V) | |
Pixel matrix: | 2448(H)×2984(V) | ||
Pixel pitch: | 143 μm | ||
Cycle time: | 6S | ||
Limiting resolution | Min. 3.7 line pair/MM | ||
A / D transition: | 16 bit | ||
Energy range | 40 - 150 kVp | ||
Maximum entrance dose (low gain) | 4 mR / frame | ||
Data output | 16-bit digital output Ethernet(1000BASE - T) | ||
Sensor Protection Board | Carbon Fiber Sheet | ||
Exposure Synchronous control: | outside | ||
power Input: | DC 24V 2A | ||
U-Arm | U-arm Vertical movement range | 500mm-1650mm | |
Focus-screen movement range | 1100mm-1800mm | ||
Rotation range | -40°- 130° | ||
Detector rotation range | -30°- 30° | ||
Collimator | Electric multi-leaf | ||
Fixed grid for bed use | Grid density: 103L/INCH, ratio:10:1 Focal distance: 110cm/180cm, 15"×18" |
Details: