Usage:
Specification | |||
1.X- ray High voltage generator | |||
1.1 | Standard output power:40kW | ||
1.2 | mA range:10-500mA | ||
1.3 | Tube voltage range:40-150kV | ||
1.4 | mAs range :0.1-250mAs | ||
1.5 | Load time range:0.001s~6.3s | ||
1.6 | Input voltage:450VDC/48VDC±10% | ||
1.7 | High voltage output master frequency:50kHz±5% | ||
1.8 | High voltage transformer structure:The power frequency | ||
2.X-ray tube | |||
2.1 | Focus range:0.6/1.2mm | ||
2.2 | Anode heat capacity :150kHu | ||
2.3 | Anode types: rotating anode, maximum speed: 3200 revolutions per minute | ||
2.4 | Pipe sleeve heat capacity: 1250kHu | ||
2.4 | The maximum voltage:150kV | ||
2.5 | The anode input power:Big focus:50kW ,Small focus:22kW | ||
2.6 | Anode angle 12° | ||
2.7 | The inherent filtration:1.3mm Al/75kV | ||
3. Collimator | |||
3.1 | The inherent filtration:1.0 mm Al /75 kV; | ||
3.2 | Collimator is equipped with LED lights(Bright 30s are automatically extinguished) | ||
3.3 | Collimator is built into measuring rulerItem | ||
4. The tube bracket structure | |||
4.1 | X-ray tube support way: Folding arm | ||
4.2 | Tube focus distance to the ground:507mm~2000mm,±5% | ||
4.3 | Arm's Folding Range :±60° | ||
4.4 | Tube Tilting Angle:±90° | ||
4.5 | Tube Vertical Rotation Angle:-30°~+90° | ||
4.6 | Collimator axial rotation :±360° | ||
5. Detector | |||
5.1 | WIFI connection ,Amorphous Silicon ,whole panel technology | ||
5.2 | Detector valid exposure size :350×430mm | ||
5.3 | Pixel size:150 µm | ||
5.4 | The effective camera size of the detector:14×17inch | ||
5.5 | Gathering matrix:≥2304×2800pixels; | ||
5.6 | Spatial resolution≥3.4lp/mm | ||
5.7 | Pixel gray level:≥14bit | ||
5.8 | Detector cooling mode: natural cooling | ||
6. Digital image workstation | |||
6.1 | CPU frequency:CORE i3 | ||
6.2 | Memory capacity:4GB | ||
6.3 | Hard drive capacity :500 GB | ||
6.4 | Display size: 17 inch LCD touch screen | ||
6.5 | Display screen Maximum resolution≥1280×1024; | ||
6.6 | 64 bit Windows operation system | ||
7. Image acquisition software | |||
7.1 | Patient registration: acquisition, find the patient data, to prepare for the photography | ||
7.2 | Radio graphic examination:setting system parameters, exposure photography and get the image | ||
7.3 | Images display, image analysis, processing and output | ||
7.4 | Data management: managing patient data, leading in, leading out , batch file, etc in accordance with DICOM3.0 | ||
7.5 | Configure calibration: the system components such as flat-panel detector, such as tube for automatic configuration and calibration | ||
7.6 | System diagnosis: monitoring system status, diagnosis of the system fault | ||
7.7 | Image Printing: Image Output to Film Printer | ||
8. Battery | |||
8.1 | Power supply way | ||
8.1.1 | Lithium battery power supply | ||
8.1.2 | The rated capacity:46.8V,10.8Ah(Lithium battery meets UN38.3 standard requirements) | ||
8.1.3 | Battery capacity: More than200 exposures, chest image dosage (exposure conditions: 120 kv, 100 mA, 20ms ) | ||
8.2 | Dual power supply, exposure under either wall current or lithium battery | ||
8.3 | Maximum charging 4 hours | ||
8.4 | Can support hand switch and remote control | ||
9. The machine parameters | |||
9.1 | Appearance size:1560 mm×650 mm×1490 mm | ||
9.2 | Whole weight (no detector):220kg | ||
10. External interface | |||
10.1 | Display interface(DVI) | ||
10.2 | Wireless network data transmission(WIFI) | ||
10.3 | Cable network interface | ||
10.4 | USB(2) | ||
11.Options | |||
11.1 | Chest rack | ||
11.2 | Moving flat bed |