Acrel Aim-M10 Hospital Insulation Monitor Groud Fault Detector Device for Unearthed AC/DC Control Systems

Min.Order: 1
Product origin: Wuxi, Jiangsu, China
Infringement complaint: complaintComplaint
US$ -1

Description
Product Description
AIM-M10 Medical Intelligent Insulation Monitor


General
Hospital IPS monitor system insulation monitor AIM-M10 with CE approval
medical intelligent insulation monitoring instrument adopts advanced microcontroller technology, which has high integration,compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection forinsulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.


Features
* real-time monitoring and fault alarming of the ground insulation resistance,
* transformer load current and transformer winding temperature of the monitored IT system,
* be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there are insulation faults,
* real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication within 2S after the fault occurs,
* relay alarm output, LED alarm indication and other faults indication functions,
* used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer management
software communications, and can monitor the operation status of IT system in real time.
* Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel toanalyze the operation conditions of system and promptly eliminate the faults.



Technical Parameters
ModelAIM-M10
VoltageAC220V( fluctuation range ±10%)
Frequency50/60Hz
Max power consumption<8VA
Insulation resistance measuring range50~999kΩ
Relative percentage accuracy0~±10%
Alarm value range10~999kΩ
Response time<2S
Measuring voltage<12V
Measuring current<42μA
Measuring range2.1~50A
Alarm value range5~50A
Measuring accuracyclass 5
DisplayDigital and percentage display
thermal resistorPT100
Measuring range-50ºC~+200ºC'
Alarm value range0ºC~+200ºC'
Output type1 channel relay output(unable to program)
Contact ratingAC 250V/3A DC30V/3A
Working temperature-10ºC~+55ºC'
Storage temperature-20ºC+70ºC'
Relative humidity5%~95%,non-condensing
Altitude≤2500m
Communicationone channel RS485,MODBUS-RTU
Rated impulse voltage/pollution class4KV/III
EMC and electromagnetic radiationIn accordance with IEC1326-2-4

Dimension(Unit:mm)







Wiring



Typical Connection



Certificate















Product Tag:
Related categories:
Scroll to Top