Touch Screen LCD Apply for Medical Patient Monitor

Min.Order: 1
Product origin: Guangzhou, Guangdong, China
Infringement complaint: complaintComplaint
US$ 300 ~ 400

Description
12.1 inch Patient Monitor Touch Screen
 

Features:
12.1" color TFT touch screen
Windows style, Multi-channel display
Big font interface
ST and arrhythmia analysis
Drug dose calculation and titration table
OxyCRG dynamic view display
Respiration monitoring with apnea alarms
Alarm review, arrhythmia review
Patient info. Input
Anti-defibrillation design
NIBP over pressure protection
Plug-in rechargeable battery, AC/DC available.
Supporting wire or wireless network.

Adopting Aiah(TM) ECG technology, anti-interference, be able to collect the weak ECG signal away from interference.
Adopting Surepeb(TM) NIBP technology, accurate measurement to high blood pressure patient and low blood pressure patient.



Technical Specifications
SP02

Nellcor(TM)  OxiMax(TM) is the world leading SpO2 technology
Nellcor(TM)  OxiMax(TM) delivers excellent performance,
even in the difficult monitoring conditions of low perfusion and signal interference, including patient motion.



ETCO2

Collaborate with Respironics, Plug and Play EtCO2 monitoring.
CAPNOSTAT mainstream sensor for optimal performance in monitoring intubated patient. Small, durable and lightweight mainstream sensor provides accurate
and reliable monitoring for all intubated patients from neonates to adults. Loflo sidestream sensor for monitoting non-intubated patient.
Flexible, compact CO2 sensor provides consistent and reliable monitoring of adult, pediatric and neonatal patients.


Multi Mounting Solutions



ECG

Lead Mode                         5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection                    I, II, III, avR, avL, avF, V
Wave form                         2 ch
Lead mode                         3 Leads ( R, L, F or RA, LA, LL)
Lead selection                    I, II, III,
Wavef orm                         1 ch
Gain                              2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV, auto

HR and Alarm Range                Adult   15 ~ 300 bpm
                                 Neo/Ped   15 ~ 350 bpm
Accuracy                          ±1% or±1bpm, which is greater
Resolution                        1 bpm
Sensitivity                       > 200 (uVp-p)
Differential Input Impedance      > 5 MΩ

CMRR                      Monitor > 105 dB
                                 Operation > 105 dB
                                 Diagnosis > 85 dB
Electrode offset potential        ±300mV
Leakage Current                   < 10 uA
Baseline Recovery                 < 3 S After Defi.
ECG Signal Range                  ±8 m V ( Vp-p )

Bandwidth                          Surgery   1 ~ 20 Hz
                                  Monitor    0.5 ~ 40 Hz
                                  Diagnostic  0.05 ~ 130 Hz
Calibration Signal                 1 (mVp-p),  Accuracy:5%
ST Segment Monitoring Range        Measure and Alarm -2.0 ~ +2.0 mV

ARR Detecting                      Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,BRADY, MISSED BEATS,
                                        PNP, PNC,Alarm Available,Review Available


Respiration

Method                            Impedance between R-F(RA-LL)
Differential Input Impedance      > 2.5 MΩ
Measuring Impedance Range         0.3~3 Ω
Base line Impedance Range         200~4 KΩ
Bandwidth                         0.1~2.5Hz
Resp. Rate                        Measuring and Alarm Range
                                  Adult 0 ~ 120 Brpm
                                  Neo/Ped 0 ~ 150 Brpm
                                  Resolution 1 Brpm
                                  Accuracy ±2 Brpm

Apnea Alarm                       10 ~ 40 S
 




NIBP

Method                           Oscillometric
Mode                             Manual, Auto, STAT                       
Measuring Interval in AUTO Mode    1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120, 180, 240,480 (Min)
Measuring Period in STAT Mode    5 Min
Pulse Rate Range           40 ~ 240 bpm
Alarm Type                  SYS, DIA, MEAN

Measuring and alarm range       Adult Mode
                                  SYS 40 ~ 270 mmHg
                                  DIA 10 ~ 215 mmHg
                                  MEAN 20 ~ 235 mmHg
                              Pediatric Mode
                                  SYS 40 ~ 200 mmHg
                                  DIA 10 ~ 150 mmHg
                                  MEAN 20 ~ 165 mmHg
                              Neonatal Mode
                                  SYS 40 ~ 135 mmHg
                                  DIA 10 ~ 100 mmHg
                                  MEAN 20 ~ 110 mmHg
Resolution                     1mmHg

Accuracy                       Maximum Mean error ±5mmHg
                              Maximum Standard deviation ±8mmHg
Overpressure Protection         Adult Mode 297±3 mmHg
                               Pediatric Mode 240±3 mmHg
                               Neonatal Mode 147±3 mmHg

SpO2
Measuring Range                 0 ~ 100 %
Alarm Range                     0 ~ 100 %
Resolution                      1 %
Accuracy                        70% ~ 100%  2 %
                               0%  ~  69% unspecified
Actualization intervalabout     1 Sec.
Alarm Delay                     10 Sec.
Pulse Rate                      Measuring and Alarm Range 20~300 bpm
                               Resolution      1bpm
                               Accuracy ±3bpm
                                 


TEMP

Channe                         2
Measuring and Alarm Range   0 ~ 50 ºC
Resolution                     0.1 ºC
Accuracy                 ±0.2 ºC
Actualization intervalabout    1 Sec.
Average Time Constant          < 10 Sec.


IBP

Label                          ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range      ART 0 ~ 300 mmHg
                              PA -6 ~ 120 mmHg
                              CVP/RAP/LAP/ICP-10 ~ 40 mmHg
                              P1 / P2-10 ~ 300 mmHg


Press Sensor                   Sensitivity 5  uV/V/mmHg
                              Impedance 300-3000Ω

Resolution                     1 mmHg
Accuracy                       ±2% or 1mmHg which is greater
Actualization interval         about 1 secretary


   

 
Standard:  ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature
Optional:  Nellcor SpO2, EtCO 2, IBP, Thermal Recorder
























Product Tag:
Related categories:
Scroll to Top