Ysav70b Hospital Mobile Medical Ambulance Respirator Ventilator

Min.Order: 1
Product origin: Guangzhou, Guangdong, China
Infringement complaint: complaintComplaint
US$ -1

Description
Hospital mobile medical ambulance respirator ventilator YSAV70B




 

The YSAV70B ventilator is gas driven by medical compressed oxygen and compressed

air. In the inspiratory phase, two streams of compressed gas (compressed oxygen and

compressed air) flow into a high performance air-oxygen mixer to form a mix of oxygen

and air with a certain pressure.

 

Maximum Minute ventilation≥18L / min
modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, SPONT
Ventilator compliance≤30 Ml/kPa
PEEP:0 ~ 1.0 kPa
Ptr:-10 ~ 10 hPa
IMV frequency:1~20 bpm
I/E Ratio:4:1~1:4, Has the inverse ratio ventilation
Tidal VolumeSetting Range: 50~1500ML     Display Range: 0~2000ml ;   resolution ratio: 10ml
Inspiratory Plateau:0~1s
Maximum Safety Pressure≤ 6.0kPa
Sigh:one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
Pressure limit range:1.0~ 6.0kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicatingLower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0 kPa
±0.2 kPa
Ventilation Volume alarm:≥ 18L/ MIN
Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
Operating conditions
Ambient temperature:10 ~ 40oC
Relative humidity:no higher than 80 %
Atmospheric pressure:860 hPa ~ 1060 hPa
Power requirement:100-120 Vac, 50/60 Hz; 
Attention: the AC power supply used for the anesthesia machine must be well     grounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.
Storage
Ambient temperature:-15oC ~ +50oC
Relative humidity:no higher than 95 %
Atmospheric pressure:86 kPa ~ 106 kPa.
 It should be stored in a room without corrosive gas and well ventilated
Package
packaging boxcomply with the requirement of GB/T 15464
 Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation
 Moisture protection and rain protection to ensure that product is protected from natural damage.
Safty & Alarm
Gas source pressureImmediate alarm; "high level alarm" indicator flickers raising an audible alarm.
Airway Pressure AlarmThe airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The "high level alarm" indicator flickers raising an audible alarm.
Tidal Volume AlarmLower than 50ml, The alarm will be raised after one breathing cycle.
Ventilation Volume AlarmLower: 0~12L/Min;   Up: 18L/ Min
Battery Voltage AlarmIt alrms when the battery voltage is lower than 10.5V.
Power On Self TestSelf test before working, assure it's working normally.
 


 