Hac6000 V Hemoglobin Hba1c Analyzer

Min.Order: 10
Product origin: Nanjing, Jiangsu, China
Infringement complaint: complaintComplaint
US$ 2540

Description

HAC6000 Full-auto Glycated Hemoglobin HbA1c Analyzer

Profile:
HbA1c Analyzer, with the methods of ion-exchange chromatographic column and single-wave visible light colorimetry, is used for analyzing HbA1c in clinical human full-blood sample. The structural characteristics and the performance parameters of the instrument conform to the actual requirement in clinical inspection. It features rapidity, correct, safety and low-pollution and has wide application in market.
 Technical Parameters of Auto Glycated Hemoglobin HbA1c Analyzer
ModelHAC6000


Basic Parameters
Test MethodChromatography/Ion-exchange liquid chromatography
Testing ItemGlycated hemoglobin HbA1c
Testing Range3%~18%
Testing ParameterPrecision(CV)≤3%
Testing Time4 minutes(Including report)
Sample TypeVenous blood(EDTA anticoagulation) peripheral finger blood


Function Parameters
Photometer415nm LED Integral flow colorimeter
Sampling ModeFull-auto 25 position sample turntable(20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position)
Testing ModeAuto Test for batch, single test for emergency
CalibrationOptional 2 points or 4 points calibration, Manual or automatic, 2 sets of high and low level calibration
Thermostatic ControlColumn and reagent thermostatic control 25±1ºC
Reagent MatchingOne piece of column an reagent consumable/300 pieces of test samples
Reagent AllocationEach standard package include A, B, C, D, eluent, Calibrator, Hemolytic agent, Pump tube


Output Parameters
DisplayLCD Display, showing real-time testing curve
Printer58mm thermal printer
Report OutputIFCC concentration value, NGSP area percentage, ADAG average glucose
Date Storage1000 test report(including testing curve)
Communication InterfaceRS232, HIS/LIS System available


Working Parameters
Power Supply220V±10%, 50Hz
Main Size(W*D*H)520*392*500mm
Sampling System Size(W*D*H)265*215*255mm
Main Weight14kg
Sampling System Weight2kg
Working EnvironmentTemperature 15~30ºC
Relative humidity≤85%







 
Scroll to Top