Specification of Ventilator
The ventilator is aerodynamic, electronic-controlled ventilator, it hasseveral respiratory modes, functions can be set throthe rerespira-tory rate, inspiratory time, tidal volume parameter to achievemanagement of the patient breathe, There are Japanese Sharp largescreeen true color LCD to show waveform and parameters, Amemerican Honeywell precision pressusure sen5 arfiow sens1SOTS,American ATMEL microprocessor, Germanic Baode proportion electromagnetism valve is Suction control valve, Japanese SMC electronic air andoxygen mixer and etc. These high quality imported parts bring outthe machine is more accurate, more reliable and more stable.
Main technical parameters and performance
Respiratory Mode:A/C, SIMV, PSV, SPONT, MMV, IPPV, SIGH, PEEP
Alarm Performances
1 Pressure upper limit (Ph) alarm: 2-6kPa, sound and light alarm
2 Pressure lower limit (PI) alarm: 0-2kPa, sound and light alarm
3Ventilation volume upper limit (MVh) alarm: more than settings,SOund and liaht alarm
SOundalar
4 Ventilation volume lower limit (MVI) alarm: less than setting,sound and light alarm
5 Power disconnection warning: immediate alarm if power.disconnection, warning time 1205.
6 Quiet sound alarm: < 2min
Parameter setting up
1 Tidal volume (Vt): 0 ~ 2000ml
2 Breathing frequency (f): 4 ~ 99 times / minute
3 SIMV Frequency: 6~ 60 timesminute
4 Inspiratory time (TD):0.2~6s
5 Breath-hold time (Tp):0to2s
6 Trigger pressure (Pt): 2kPa~ 2kPa,
7 PEEP pressure:0.1 ~ 2kPa,
8 PSV pressure:0~5 kPa
9 Safety largest gas pressure: <6 kPa
10Breathing Oxygen Concentration; 40-100%,Consecutive adjustable
Overall performance
1 Drive mode: aerodynamic, electronic-controlled 10.4 inch colorful TFT Screen
2 Aerodynamic source pressure: 0.28~0.6MPa
3 Mainframe input power: 65VA.
4 Complete machine noise: <65 dB (A).
5 Compliance system: <4 m/100Pa