Electrical Connection |
The power supply voltage: |
(4~20)mA/HART (Two wire system) The power supply and the output current signal sharing a two core shield cable. The supply voltage range see technical data. For intrinsically safe type must be a safety barrier between the power supply and the instrument. |
(4~20)mA/HART(Four wire system) Separate power supply and the current signal, respectively using a two-core shielded cable. The supply voltage range see technical data. |
RS485 / Modbus Power supply and Modbus signal line separated respectively using a two-core shielded cable, the power supply voltage range see technical data. |
Connection mode: |
Technical Parameters |
The outer shell The seal between the shell and the shell cover Silicone rubber Casing window Polycarbonate The ground terminal Stainless steel |
The power supply voltage Two wire system The standard type (16 ~ 26) V DC Intrinsically safe (21.6 ~ 26.4) V DC Power dissipation max 22.5mA / 1W Allowable ripple - <100Hz Uss<lV - (100~100K) Hz Uss<l0mV |
The cable parameters Cable entrance / plug 1 M20xl.5 cable entrance 1 blind plug Terminal Conductor cross section 2.5mm² |
Output parameters The output signal (4 ~ 20) mA/RS485 Communication protocol HART Resolution 1.6μA Fault signal Constant current output; 20. 5mA 22mA 3.9mA The integral time (0 ~ 36) s, adjustable |
Blind area the ends of the antenna |
The maximum distance measurement 70 meters |
Microwave frequency 26GHz |
Communication interface HART communication protocol |
The measurement interval about 1 second (depending on the parameter settings) |
Adjust the time about 1 second (depending on the parameter settings) |
Display resolution 1 mm |
Working storage and transportation temperature (-40~100) ºC |
Process temperature (the temperature of the antenna part) (-40~250)ºC |
Pressure Max.4MPa |
Seismic Mechanical vibration l0m/s², (10 ~ 150) Hz |
Meter Linearity |
Emission angle Depending on the size of the antenna -¢76mm 12° -¢96mm 8° -¢121mm 6° Precision See chart |