Function
Continuous/random test mode, emergency sample priority test. Available to test according to sample sequence or test item sequence;
Test methods: End-point, Fixed-time, Kinetic; supportive for dual-reagent and di-wavelength test;
The reagent probe has liquid level detection function and full collision protection function;
Automatic cleaning function of the reaction cup;
The external barcode scan module can be connected, the seamless connection with Laboratory Information Management System(LIS) is supported, which achieves real-time data sharing;
Recalculation function, user can select different conditions as necessary to recalculate results;
With print preview and print function;
Provide three kinds of quality control charts: multiple-rule quality control chart, cumulative quality control chart and twin plot quality control chart;
Provide statistical reports, including test statistics, workload statistics, expense statistics, and results statistics;
Operations of add, modify or delete items are supportive;
Semiconductor cooling system, uninterrupted continuous cooling, stable and reliable.
Continuous/random test mode, emergency sample priority test. Available to test according to sample sequence or test item sequence;
Test methods: End-point, Fixed-time, Kinetic; supportive for dual-reagent and di-wavelength test;
The reagent probe has liquid level detection function and full collision protection function;
Automatic cleaning function of the reaction cup;
The external barcode scan module can be connected, the seamless connection with Laboratory Information Management System(LIS) is supported, which achieves real-time data sharing;
Recalculation function, user can select different conditions as necessary to recalculate results;
With print preview and print function;
Provide three kinds of quality control charts: multiple-rule quality control chart, cumulative quality control chart and twin plot quality control chart;
Provide statistical reports, including test statistics, workload statistics, expense statistics, and results statistics;
Operations of add, modify or delete items are supportive;
Semiconductor cooling system, uninterrupted continuous cooling, stable and reliable.
Performance
Light source: Halogen-tungsten lamp, 12V 20W;
Temperature control: 37°C;
Temperature accuracy: ±0.1°C;
Reaction disc: 81 cuvettes, optical length 5mm;
Cuvette: Semi-permanent plastic cuvette, automatic cleaning, recycling;
Wavelength: Standard 8 wavelengths: 340nm-670nm; customized wavelength available;
Stray light: absorbance ≥2.5;
Absorbance linearity: The maximum absorbance of relative deviation within ±2% is no less than 4.5;
Absorbance accuracy: The error is no more than ±0.015 when the absorbance is 0.5, and the error is no more than ±0.025 when the absorbance is 1.0;
Absorbance stability: The change of absorbance is no more than 0.005;
Absorbance repeatability: CV≤1.0%;
Temperature accuracy and fluctuation: temperature value within 37°C±0.2°C, the fluctuation is no more than ±0.1°C;
Sample carry-over rate: ≤0.5%;
Sampling accuracy and CV: Accuracy≤±5%, CV≤2%;
Within-run precision (CV%): ALT≤5%; TP≤2.5%; UREA≤2.5%;
Data storage: automatic storage, automatic backup, permanent storage of various types of data;
Power supply: AC100V-240V, 50/60Hz.
Light source: Halogen-tungsten lamp, 12V 20W;
Temperature control: 37°C;
Temperature accuracy: ±0.1°C;
Reaction disc: 81 cuvettes, optical length 5mm;
Cuvette: Semi-permanent plastic cuvette, automatic cleaning, recycling;
Wavelength: Standard 8 wavelengths: 340nm-670nm; customized wavelength available;
Stray light: absorbance ≥2.5;
Absorbance linearity: The maximum absorbance of relative deviation within ±2% is no less than 4.5;
Absorbance accuracy: The error is no more than ±0.015 when the absorbance is 0.5, and the error is no more than ±0.025 when the absorbance is 1.0;
Absorbance stability: The change of absorbance is no more than 0.005;
Absorbance repeatability: CV≤1.0%;
Temperature accuracy and fluctuation: temperature value within 37°C±0.2°C, the fluctuation is no more than ±0.1°C;
Sample carry-over rate: ≤0.5%;
Sampling accuracy and CV: Accuracy≤±5%, CV≤2%;
Within-run precision (CV%): ALT≤5%; TP≤2.5%; UREA≤2.5%;
Data storage: automatic storage, automatic backup, permanent storage of various types of data;
Power supply: AC100V-240V, 50/60Hz.
Accessories
Software CD
User manual
Internal hexagonal wrench
National standard power cord
Grounding cable
Network cable
Pricker
Cuvette
Concentrated cleaning solution
R1/R2 plastic bottle
Bucket assembly (for pure water, waste liquid, cleaning agent respectively)
Software CD
User manual
Internal hexagonal wrench
National standard power cord
Grounding cable
Network cable
Pricker
Cuvette
Concentrated cleaning solution
R1/R2 plastic bottle
Bucket assembly (for pure water, waste liquid, cleaning agent respectively)
Physical characteristic
Dimension of main unit: 880mm(L)×650mm(W)×550mm(H)
Weight of main unit: 105kg
Dimension of cabinet: 880mm(L)×650mm(W)×630mm(H)
Weight of cabinet: 55kg
Operating Environment:
Temperature: 10~30ºC
Relative humidity: ≤85%
Atmospheric pressure: 86kPa~106kPa
Storage Environment:
Temperature: -10~55ºC
Relative humidity: ≤93%
Atmospheric pressure: 86kPa~106kPa
Dimension of main unit: 880mm(L)×650mm(W)×550mm(H)
Weight of main unit: 105kg
Dimension of cabinet: 880mm(L)×650mm(W)×630mm(H)
Weight of cabinet: 55kg
Operating Environment:
Temperature: 10~30ºC
Relative humidity: ≤85%
Atmospheric pressure: 86kPa~106kPa
Storage Environment:
Temperature: -10~55ºC
Relative humidity: ≤93%
Atmospheric pressure: 86kPa~106kPa