Description: | This density tester is based on GB/T5162-2006/ISO3953:1993(tap density measuring of metal powder). For non-metallic powder,This instrument extended seveal functions.such as"vibration amplitude" from 3m standard extended to 1mm-15mm;"vibration frequency"from 100-300 times per minute extended to 0-300 times per minute."vibration times"from 3000 times standard extended to 0-99999 times.All of them can be setup freely. Powder density tester consist of adjustable-speed motor,vibration assembly,micro-computer and micro printer.,etc.It has the feature of compact structure,stable,and easy operation.,etc. | |||||
Standards: | GB/T5162-2006/ISO3953:1993(tap density measuring of metal powder) | |||||
Characters: | 1 Precision: Automatically keep five significant figures, accurate to four decimal places. | |||||
2 Intelligent judge: this equipment can be smart judge when entering the required information. Decimal point is automatically shielded | ||||||
when inputting tap number and during which period the vibration is halted. The results show the actual number of vibrations, the experiment is authentic. | ||||||
3 Easy operation: No need to adjust to an integer. | ||||||
Specifications: | Weight of testing object:<=500.00g | |||||
Volume of testing object:<=250.00ml(standard accessory including 25ml,50ml,100ml,250ml measuring cylinder) | ||||||
Single vibration times:<=0-99999 times(the result will come out as bulk density when input zero times) | ||||||
Allowable motor force:0.86 N.m. | ||||||
Vibration frequency:maximum 300 round per minute(0-300 round per minute is adjustable,USA pharmacopeia standard is 300 round per mintute) | ||||||
Vibration amplitude:maximum 15mm(1mm-15mm is adjustable,International standard is mm) | ||||||
Repeatability error:<=1% | ||||||
Accuracy error:<=1% |