Microprocessor based servo controlled temperature system
Three modes of control: Pre-warm, manual and automatic
Set temperature and central skin temperature, peripheral temperaqture,
air temperature,APGAR timer,heating power display separately
8" LCD colorful touch screen
Carbon steel heater for more effectivity
Special design Inclined heater head is more convenience for X ray test
Digital skin temperature sensor, the data is accurate and stable
Alloyed bracket, fastness
Alarms functions for safety includes power failure, temperature deviation,temperature probe failure, over temperature
Integral timer to record therapy time
Bigger mattress,
Skin temperature trend for 5min
Electric infant bed tilt angle is ± 12º
Transparent protector folded down,with two infusion seals
Convenient X-ray cassette tray under infant bed
Electric height adjusted suit for operation
With observe light, oxygen supply system and LED phototherapy
Optional Parts: low pressure aspirator ,internal / external
infant resuscitator (T piece) , Air and oxygen blender ,
Gel mattress
Main technical parameter
1. Power supply: ~220V±22V / 50Hz±1Hz
2. Power input: 1100VA
3. Normal working condition:
a. Environment temperature: 18ºC~30ºC;
b. Relative humidity: 30%~75%;
c. Atmospheric pressure: 700~1060hPa;
4. Temperature control range: 32ºC~38ºC
5. Precision of temperature control: ≤0.5ºC
6. Functional alarm:
6.1. Over-temp: 39ºC
The secondary over temp to shut off power: 40ºC(cut off the heating power, alarm with sound and light)
6.2. Deviation alarm: When the temp goes steady and the deviation reaches ±1ºC(+1ºC, the heating power will be shut off), alarm with sound and light.
6.3. Sensor alarm: When the skin temp sensor is in short circuit or open circuit state, shut off the heating power, alarm with sound and light; the sensor overruns the radiation area and temp deviation reaches to -1ºC for about one minute, the machine will alarm with sound and light automatically and keep certain power to heat up.
6.4. Power failure alarm: When the power breaks off, alarm with sound and light.
6.5. Bed temp uniformity: ≤2ºC
6.6. Skin temp sensor precision: ≤0.3ºC
6.7. Increasing temp time: 45'
6.8. Transportation and storage
a. Environment temp: -40ºC~+55ºC
b. Relative humidity range: ≤95%
c. Atmospheric pressure range: 500~1060hPa
1.Who we are?
We are based in Jiangsu, China, near Shanghai,start from 2008.We are the earliest manufacturer of export funeral products in China,sell to South America(20.00%),Southeast Asia(20.00%),North America(10.00%),Africa(10.00%),Eastern Asia(10.00%),Western Europe(10.00%),Southern Europe(10.00%),Domestic Market(10.00%).