MCI0039 portable medical diagnostic high-frequency X-ray machine for the protection of
Main Technical Parameters
a) Power Conditions
The power supply voltage is single-phase AC 220V, 50Hz, sine wave. The power
supply voltage range is 220V ± 10%, and the frequency value range is 50Hz ± 1Hz.
The internal resistance of the power supply should not be greater than 1.0 Ω, and
the power supply capacity should not be less than 6.5 KVA.
Input Power:8.0kVA; Over-current Releaser:20A/220V; Fuse:20A/220V;
b) Working Environment Conditions
1) Ambient Temperature:5ºC~40ºC;
2) relative humidity:30%~75%;
3) Atmospheric Pressure:70.0kPa~106.0kPa.
c) Working Frequency
The working frequency is not less than 30 kHz in the photographic state.
d) Standard Output Power
Intermittent Mode:
The maximum output power 5.0kW(100kV,50mA,0.1s);
The maximum voltage tube is 110kV,and the photography maximum currency tube is
40m;
The maximum tube current of the photography is 100mA, and the corresponding
maximum tube voltage is 50kV.
e) Exposure Time
The exposure time is set by digital setting and grading adjustment, and the adjustment
range is from 0.01s to 6.3ss;
f) Adjusting Tube Voltage and Current
aggregate capacity | 5KW |
service voltage | 220V |
mA | 32-100 mA |
KV | 40-110 KV step 1 KV |
exposure time | 0.01~6.3s |
Speed and heat capacity | 42KHU |
Generator frequency | 30Khz |
Imaging area | 14*17 inch |
Pixel size | 140um |
A/D conversion | 16bits |