- Weight: 2.8kg(excluding accessories) Power Supply
- Adapter: Output 9VDC, 5A
- Voltage:AC100~240V, 50/60HZ Display
- 10.4" color TFT LED (0-135° rotation)
- Resolution:800*600 pixels Battery (Built-in)
- Type: Rechargeable lithium battery
- Charge Cycle: ≥500 times
- Working time: 2 hours (optional larger battery for 4-5 hours)
Recorder (Built-in)
- Method: Thermal printer
- Paper: Z-fold, thermosensitive
- Paper width: 112 mm (4.41 in)
- Printing speed: 1/2/3cm/min
- Trace: Max. 4 tracks
- Recording way: Real-time Recording,Review Printing Alarm
- Level: Low, medium and high
- Indication: Auditory, Visual, Delay of FHR Alarm Input Device
- Touch screen (option)
- Keypad input (Standard)
- Mouse/ Keyboard input (option)
- USB port for patient data/NST report backup(option)
System Output & Extensible Interface
- Ethernet Network: 1 standard RJ45 socket
- USB Port:1 Trend & Reviewing
- Cardiotocography (CTG): Waveform Review: 8640 hours
- Vital Signs: Trend: 2160 hours
Waveform Review: 2 hours
NIBP Review: 2000 groups
- Alarm Review: 2000 groups
Ultrasound (FHR)
- Transducer type: 12-Crystal ultrasound transducer
- Measure method: Wide beam pulse doppler
- FHR Measurement range: 30 to 240 bpm (US Standard) ,50 to 210 bpm (Europe Standard)
- Accuracy: ±1 bpm
- Intensity: <3 mW/cm2
- Ultrasound frequency: 1MHz±3%
- Resolution: 1 bpm
TOCO
- Measurement method: Strain gauge sensor element
- Measure range: 0 ~ 100 units
- Resolution: 1 unit
- Nonlinear deviation : ≤±8%
- Baseline setting: 0,5,10,15, 20 can be selected
- Auto offset correction: 3 second after connecting the transducer
- Auto zero adjust: TOCO value is set to zero negativefollowing a negative Measure value for 5 seconds