TD4200 Three-phase and DC Meters Test Equipment

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 90800 ~ 155400

Description
TD4200 Testing DeviceDevice for Three-phase and DC Meters

1.Summary
TD4200 is a multi-function comprehensive test-bed for the calibration of electrical measuring instruments. It integrates precision 3PH standard source, high-precision 3PH standard meter, precision DC standard source, test-bed, computer and special software. Can be used to calibrate three-phase AC/DC meter, AC standard source, AC/DC transmitter, etc.

2. Features

  1. Accuracy class: 0.01, 0.02;
  2. 3PH voltage output:0.3 V~825 V;
  3. 3PH current output:0.3 mA~110 A;
  4. 3PH standard meter (option): Tunkia TD33 series three-phase multi-function standard meter; (Note: also support COM3003,RD33,K2006, etc.)
  5. DC power output (option): 1100 V / 33A or 110A;
  6. Frequency and phase adjustable;
  7. Harmonic function / AC energy meter test function;
  8. Test software.

3. Applications



 
 Calibrate Meters or Sources
  1. 3PH voltage output: 0.3 V~825 V @ 45 Hz~70 Hz.
  2. 3PH current output: 0.3 mA~110 A @ 45 Hz~70 Hz.
  3. 3PH power output: voltage and current combination.
  4. DC voltage output: 5 mV~1100 V (option).
  5. DC current output: 10 μA~110 A (option).
  6. DC power output (option): voltage and current combination.
  7. Applied to the calibration of AC/DC voltage and current meter, active power meter, DC power meter, reactive power meter, volt-ampere meter and power factor meter.
  8. TD33 series standard meter (option): the best accuracy of measurement for 3PH power and electrical energy could reach 100 ppm; Applied to the calibration of single-phase or three-phase power standard source.


 

4. Characteristics

 Wide Output
  1. It can meet most single-phase / three-phase meters or DC meters. 
 
 Wide Input (TD33 series option)
  1. It can meet most single-phase / three-phase meters or DC source. 


 
 Multiple Output Mode
  1. "Direct output"mode,User can set output value by physical key or touch screen.
  2. Three-phase unit adjustment or single phase adjustment.
  1. "Rotary Knob"mode,User can setting in clockwise direction or anticlockwise direction.


 
 Multiple Output Mode
 
  1. Touch "Calibration point"of screen for "% setting".


 
 Testing Software(Option)
  1. RS232 communication interface,software customizable.

5. Specifications

5.1 Three-Phase Power Source

5.1.1 Three-Phase Voltage Output

Voltage RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (mA)
Class 0.02Class 0.01Class 0.02Class 0.01
3 V0.01 mV0.0050.003120 + 8060 + 40500 mA
10 V0.1 mV0.0050.003120 + 8060 + 40500 mA
30 V0.1 mV0.0050.003120 + 8060 + 40500 mA
100 V0.1 mV0.0050.003120 + 8060 + 40350 mA
300 V1 mV0.0050.003120 + 8060 + 40120 mA
600 V1 mV0.0050.003120 + 8060 + 4050 mA
750 V10 mV0.0050.003120 + 8060 + 4040 mA
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).
  1. Output range: 0.3 V~825 V, distortion: < 0.2%;
  2. Short circuit and overload protection.

5.1.2 Three-Phase Current Output

Current RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (V)
Class 0.02Class 0.01Class 0.02Class 0.01
3 mA0.01 μA0.0050.003120 + 8060 + 4050 V
10 mA0.1 μA0.0050.003120 + 8060 + 4050 V
30 mA0.1 μA0.0050.003120 + 8060 + 4050 V
100 mA1 μA0.0050.003120 + 8060 + 4050 V
300 mA1 μA0.0050.003120 + 8060 + 4050 V
1 A10 μA0.0050.003120 + 8060 + 4050 V
3 A10 μA0.0050.003120 + 8060 + 4030 V
10 A0.1 mA0.0050.003120 + 8060 + 406 V
30 A0.1 mA0.0050.003120 + 8060 + 402.5 V
100 A1 mA0.0050.003120 + 8060 + 401.2 V
  1. Output range: 0.3 mA~110 A, distortion: < 0.2%;
  2. Open circuit and overload protection.

5.1.3 Frequency/Phase/Harmonic

SymmetryVoltage superior to 0.2%; Current superior to 0.5%; Phase superior to 0.5°
FrequencyRange:45 Hz~70 Hz;
Adjustment fineness:0.001 Hz;
Accuracy:±0.01Hz (class 0.02), ±0.005Hz (class 0.01).
PhaseRange:0.000 0°~359.999 9°;
Adjustment fineness:0.001°;
Accuracy:±0.01° (class 0.02), ±0.005° (class 0.01).
Harmonic2nd~21st harmonic;
Amplitude 0~25% adjustable;
Phase 0~359.99°adjustable

5.1.4 Three-Phase Power Output

TypeCurrent rangeStability
( %/min )
Accuracy
( ± %*FS)[2]
Class 0.02Class 0.01Class 0.02Class 0.01
active power |cosφ| ≥ 0.5100 mA…100 A0.010.0050.020.01
10 mA…50 mA0.010.0050.050.02
reactive power   |sinφ| ≥ 0.5100 mA…100 A0.010.0050.050.02
10 mA…50 mA0.020.010.10.05
apparent power100 mA…100 A0.010.0050.050.02
10 mA…50 mA0.020.010.10.05
power factor100 mA…100 A0.010.0050.050.02
10 mA…50 mA0.020.010.10.05
Note [2] :FS= voltage range ×current range
  1. Range of power factor: -1.000 000…0.000 000…1.000 000.

5.2 Three-Phase Standard Meter (option, TD33 series)

5.2.1 Three-Phase Voltage Input

RangeResolutionAccuracy
±(ppm of reading + ppm of range) [1]
Temperature Coefficient       @ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02Class 0.01Class 0.02Class 0.01
60 V10 μV60 + 4030 + 200.5 + 0.50.25 + 0.25
120 V0.1mV60 + 4030 + 200.5 + 0.50.25 + 0.25
240 V0.1mV60 + 4030 + 200.5 + 0.50.25 + 0.25
480 V0.1mV60 + 4030 + 200.5 + 0.50.25 + 0.25
  1. Measuring range: 6 V~528 V (support customized wider range), 7 digit display, manual / automatic range.
 

5.2.2 Three-Phase Current Input

RangeResolutionAccuracy
±(ppm of reading + ppm of range) [1]
Temperature Coefficient       @ (15~30)°C
(ppm*RD+ppm*RG) /°C
Class 0.02Class 0.01Class 0.02Class 0.01
5 mA1 nA240 + 160120 + 8010 + 105 + 5
10 mA10 nA120 + 8060 + 405 + 53 + 3
20 mA10 nA120 + 8060 + 401.25 + 1.250.75 + 0.75
50 mA10 nA60 + 4030 + 200.5 + 0.50.25 + 0.25
100 mA0.1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
200 mA0.1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
500 mA0.1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
1 A1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
2 A1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
5 A1 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
10 A10 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
20 A10 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
50 A10 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
100 A100 μA60 + 4030 + 200.5 + 0.50.25 + 0.25
  1. Measuring range: 0.2 mA~120 A, 7 digit display, manual / automatic range.
 

5.2.3 Frequency / Phase

TypeClass 0.02Class 0.01
FrequencyRange45 Hz~65 Hz45 Hz~65 Hz
Resolution0.000 01 Hz0.000 01 Hz
Accuracy± 0.005%*RD± 0.005%*RD
PhaseRange0~360°(I ≥ 50mA)0~360°(I ≥ 50mA)
Resolution0.000 1°0.000 1°
Accuracy± 0.006°± 0.003°
 

5.2.4 Three-Phase Power Input

Voltage RangeCurrent RangePower FactorAccuracy
Class 0.02Class 0.01
30 V ≤ U ≤ 480 V50 mA ≤ I ≤120 A0.5L~1~0.5C± 0.02%*reading± 0.01%* reading
10 mA ≤ I < 50 mA1± 0.02%* reading± 0.01%* reading
0.5L~1~0.5C± 0.04%* reading± 0.02%* reading
3 mA ≤ I < 10 mA1± 0.04%* reading± 0.02%* reading
0.5L~1~0.5C± 0.08%* reading ± 0.04%* reading
0.2 mA ≤ I < 3 mA1± 0.04%*reading
×3mA/I
± 0.02%*reading
×3mA/I
  1. Power / electrical energy input range: combine AC voltage range and AC current range;
  2. Power factor input range: -1.000 000…0.000 000…1.000 000;
  3. Output pulse of standard electrical energy: at full scale, the high frequency correspond to 60 kHz and low frequency correspond to 6 Hz;
  4. Input pulse of standard electrical energy: frequency≤ 200 kHz, voltage: 0…3.3 V…24 V.
 

5.3 DC Power Source (option)

5.3.1 DC Voltage Output

Voltage RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (mA)
Class 0.02Class 0.01Class 0.02Class 0.01
10 mV0.1 μV0.0050.003100 + 5 μV100 + 5 μV100 mA
30 mV0.1 μV0.0050.003100 + 5 μV100 + 5 μV100 mA
100 mV1 μV0.0050.003100 + 5 μV100 + 5 μV100 mA
300 mV1 μV0.0050.003100 + 5 μV100 + 5 μV100 mA
1 V10 μV0.0050.002100 + 5060 + 40500 mA
3 V10 μV0.0050.002100 + 5060 + 40500 mA
10 V0.1 mV0.0050.002100 + 5060 + 40500 mA
30 V0.1 mV0.0050.002120 + 8060 + 40800 mA
100 V1 mV0.0050.002120 + 8060 + 40300 mA
300 V1 mV0.0050.002120 + 8060 + 40100 mA
600 V10 mV0.0050.002120 + 8060 + 4050 mA
1000 V10 mV0.0050.002120 + 8060 + 4030 mA
  1. DC voltage output: 1 mV~1100 V, ripple ratio: < 1%;
  2. Short circuit and overload protection.
 

5.3.2 DC Current Output

Current RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (V)
Class 0.02Class 0.01Class 0.02Class 0.01
10 μA0.1 nA0.010.01300 + 200300 + 20011 V
30 μA0.1 nA0.0050.005120 + 80120 + 8011 V
100 μA1 nA0.0050.005120 + 80120 + 8011 V
300 μA1 nA0.0050.005120 + 80120 + 8011 V
1 mA10 nA0.0050.003120 + 80100 + 5011 V
3 mA10 nA0.0050.003120 + 80100 + 5011 V
10 mA0.1 μA0.0050.003120 + 80100 + 5011 V
30 mA0.1 μA0.0050.003120 + 80100 + 5011 V
100 mA1 μA0.0050.003120 + 80100 + 5011 V
300 mA1 μA0.0050.002120 + 8060 + 4011 V
1 A10 μA0.0050.002120 + 8060 + 4011 V
3 A10 μA0.0050.002120 + 8060 + 404 V
10 A100 μA0.0050.002120 + 8060 + 404 V
30 A3100 μA0.0050.002120 + 8060 + 403 V
100 A31 mA0.0050.002120 + 8060 + 403 V
  1. DC current output: 1 μA~33 A or 100 A optional, ripple ratio: < 1%;
  2. Open circuit and overload protection.
 

5.3.3 DC Power Output

DC power outputRangeCombine DC voltage and current range
AccuracyAddition of voltage and current accuracy.
 

5.3.4 DC Meter (Transmitter)

RangeMeasuring RangeAccuracyRipple Measuring RangeRipple Accuracy
1 V± ( 0~1.2 ) V± 0.01%*range0~30 mV± 1 mV
10 V± ( 0~12 ) V± 0.01%* range0~300 mV± 10 mV
2 mA±( 0~2.4) mA± 0.01%* range0~60 μA± 2 μA
20 mA± ( 0~24 ) mA± 0.01%* range0~600 μA± 20 μA
  1. Response time: range: 0~1000 ms, accuracy: ±40 ms.

6. General Specifications

Power SupplyAC ( 220 ± 22 ) V,( 50 ± 2 ) Hz
Temperature PerformanceWorking temperature:0°C~45°C;
Storage temperature:-20°C~70°C
Humidity PerformanceWorking humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C
Storage humidity:(20%~80%) R·H,non-condensing
InterfaceRS232
 

7. Ordering Information

Company Profile
 



Product Tag:
Related categories:
Scroll to Top