TD4100 Testing System for Three-phase and DC Meters

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 26300 ~ 29700

Description
TD4100 Testing System for Three-phase and DC Meters

1. Summary

TD4100 is a multifunctional 3PH power source. It integrates AC/DC power output function, harmonic function, transmitter test function, AC energy meter test function. Can be used to calibrate AC/DC voltmeter, AC/DC ammeter, AC/DC power meter, electric energy meter and transmitter, etc.
 

2. Features

  1. 3PH AC voltage output:1 V~825 V
  2. 3PH AC current output:2 mA~110 A
  3. Frequency:45 Hz~70 Hz
  4. U-I phase:~360°
  5. DC voltage output:5 mV~1100 V
  6. DC current output:1 μA~33 A
  7. Accuracy class: 0.02, 0.05.
  8. 2nd~21st harmonic output (option)
  9. Transmitter test function (option)
  10. AC energy meter test function(option)
  11. Test software(option)

3. Application



 

4. Characteristics

Wide Output
  1. It can meet most single-phase / three-phase meters or DC meters.



 
Multiple Output Mode
  1. "Direct output"mode,User can set output value by physical key or touch screen.
  2. Three-phase unit adjustment or single phase adjustment.
  1. "Remote Control Box(option)",by operation of coarse tuning-fine tuning,can calibrate analog meter.


 
Multiple Output Mode
  1. Touch "Calibration point"of screen for "% setting".
  1. "Rotary Knob"mode,User can setting in clockwise direction or anticlockwise direction.


 

Harmonics Function
S/NFunction declaration
1Set amplitude of harmonic.
2Set phase of harmonic(fundamental wave).
3Choose 2nd~21st harmonics channel output.
4Oscilloscope function, user can observe voltage and current waves.
5Display frequency spectrum of harmonic by histogram.(fundament wave is 100%)
 

5. Specifications

5.1 Three-Phase Voltage Output

Voltage RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (mA)
Class 0.05Class 0.02Class 0.05Class 0.02
15 V0.1 mV0.010.005300 + 200120 + 80500
57.7 V0.1 mV0.010.005300 + 200120 + 80500
100 V1 mV0.010.005300 + 200120 + 80300
220 V1 mV0.010.005300 + 200120 + 80140
380 V1 mV0.010.005300 + 200120 + 8080
600 V1 mV0.010.005300 + 200120 + 8050
750 V10 mV0.010.005300 + 200120 + 8030
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%).
  1. Three-phase voltage output:1 V~825 V;
  2. Degree of distortion:<0.2%
  3. Short circuit and overload protection
 

5.2 Three-Phase Current Output

Current RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (V)
Class 0.05Class 0.02Class 0.05Class 0.02
20 mA0.1 μA0.010.005300 + 200120 + 8060
50 mA0.1 μA0.010.005300 + 200120 + 8060
100 mA1 μA0.010.005300 + 200120 + 8060
200 mA1 μA0.010.005300 + 200120 + 8030
500 mA1 μA0.010.005300 + 200120 + 8030
1 A10 μA0.010.005300 + 200120 + 8030
2 A10 μA0.010.005300 + 200120 + 806
5 A10 μA0.010.005300 + 200120 + 806
10 A100 μA0.010.005300 + 200120 + 802.5
20 A100 μA0.010.005300 + 200120 + 801.2
50 A100 μA0.010.005300 + 200120 + 800.8
100 A1 mA0.010.005300 + 200120 + 800.8
  1. Three-phase current output:2 mA~110 A;
  2. Degree of distortion:<0.2%
  3. Current open-circuit and overload protection
 

5.3 Frequency / Phase / Harmonic

SymmetryVoltage superior to 0.2%;Current superior to 0.5%;Phase superior to 0.5°
FrequencyRange:45 Hz~70 Hz;Adjustment fineness:0.001 Hz;
Accuracy:±0.02 Hz (class 0.05), ±0.01 Hz(class 0.02)
PhaseRange:0.000 0°~359.999 9°;Adjustment fineness:0.001°;
Accuracy:±0.02° (class 0.05), ±0.01°(class 0.02)
Harmonic(optional)2nd~21st harmonic;
Amplitude 0~25% adjustable;Phase 0~359.99°adjustable
 

5.4 Three-Phase Power Output

Current RangeStability
( %/min )
Accuracy
( ± %*FS)[2]
Class 0.05Class 0.02Class 0.05Class 0.02
Active power |cosφ|≥0.50.010.0050.050.02
Reactive power |sinφ|≥0.50.020.010.10.05
Apparent power0.020.010.10.05
Power factor0.020.010.10.05
Note [2] :FS= voltage range ×current range
  1. Power factor range:-1.000 000…0.000 000…1.000 000
 

5.5 DC Voltage Output

Voltage RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (mA)
Class 0.05Class 0.02Class 0.05Class 0.02
75 mV0.1 μV0.010.005300 + 200120 + 12010
100 mV1 μV0.010.005300 + 200120 + 8010
300 mV1 μV0.0050.005300 + 200120 + 8010
1 V10 μV0.0050.005300 + 200120 + 8010
3 V10 μV0.0050.005300 + 200120 + 8010
10 V100 μV0.0050.005300 + 200120 + 8010
30 V100 μV0.0050.005300 + 200120 + 80500
60 V100 μV0.0050.005300 + 200120 + 80150
100 V1 mV0.0050.005300 + 200120 + 80150
300 V1 mV0.0050.005300 + 200120 + 8050
600 V1 mV0.0050.005300 + 200120 + 8025
1000 V10 mV0.0050.005300 + 200120 + 8015
  1. DC voltage output range:5 mV~1100 V,
  2. Ripple factor:< 1%
  3. Short circuit and overload protection
 

5.6 DC Current Output

Current RangeResolutionStability
( %/min )
Accuracy
±(ppm of reading + ppm of range) [1]
Max
Burden (V)
Class 0.05Class 0.02Class 0.05Class 0.02
10 μA0.1 nA0.010.005300 + 200120 + 8010
30 μA0.1 nA0.010.005300 + 200120 + 8010
100 μA1 nA0.010.005300 + 200120 + 8010
300 μA1 nA0.010.005300 + 200120 + 8010
1 mA10 nA0.010.005300 + 200120 + 8010
3 mA10 nA0.010.005300 + 200120 + 8010
10 mA100 nA0.010.005300 + 200120 + 8010
30 mA100 nA0.010.005300 + 200120 + 8010
100 mA1 μA0.010.005300 + 200120 + 8010
300 mA1 μA0.010.005300 + 200120 + 804
1 A10 μA0.010.005300 + 200120 + 804
3 A10 μA0.010.005300 + 200120 + 804
10 A100 μA0.010.005300 + 200120 + 803
30 A100 μA0.010.005300 + 200120 + 802.5
  1. DC current output range:1 μA~33 A,
  2. Ripple factor:< 1%
  3. Current open-circuit and overload protection
 

5.7 DC Meter(option)

RangeMeasurement RangeAccuracyMeasurement Range of RippleAccuracy of Ripple
1 V± ( 0~1.2 ) V± 0.01%*range0~30 mV± 1 mV
10 V± ( 0~12 ) V± 0.01%* range0~300 mV± 10 mV
2 mA±( 0~2.4) mA± 0.01%* range0~60 μA± 2 μA
20 mA± ( 0~24 ) mA± 0.01%* range0~600 μA± 20 μA
Note:This DC Meter is used to test second signal of Transmitter
 

5.8 AC Energy Testing(option)

TypeAccuracy
Class 0.05Class 0.02
Active electrical energy± 0.1%*reading± 0.05%*reading
Reactive electrical energy± 0.2%*reading± 0.1%*reading
  1. Energy pulse output:full range for 60 kHz
  2. Energy pulse input:max frequency is 200 kHz,pulse level:3 V~12 V
  3. Constant setting of Electrical energy meter:1…1000000 imp./kwh or 1…1000000 imp./ws

6. General Specifications

Power SupplyAC ( 220 ± 22 ) V,( 50 ± 2 ) Hz
Temperature PerformanceWorking temperature:0°C~45°C;
Storage temperature:-20°C~70°C
Humidity PerformanceWorking humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C
Storage humidity:(20%~80%) R·H,actnon-condensing
InterfaceRS232
 