TH0210 Standard Resistance Measuring Apparatus/Instrument

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 48300 ~ 51000

Description
TH0210 Standard Resistance Measuring Apparatus


1. Summary
TH0210 is a direct-current-comparator bridge designed for the automatic measurement and display of the ratio of two resistances to high accuracy. It's applied for the measurement, comparison, transmission, verification of standard resistance.

2. Features
Current output: 100 μA ~ 3.2 A. 
Resistance measurement: 1 mΩ ~ 100 kΩ(customizable).
0.5 ppm ~ 3 ppm(External Primary Standard Resistor)
Typical value of test period: about 4.5 minutes.
RS232 and LAN interfaces.
LCD touch screen.
Statistical analysis.
Dedicated software(optional).
 
3. Applications
Calibrate 2-wire/4-wire resistor
Recommend 4-wire mode when the value of resistor is ≤ 10 kΩ.

Establish automatic resistance measurement system
Multiple (up to 10) standard resistors.
The channel of reference standard resistance can be automatically switched according to the value of measured resistance.

  • Statistical analysis
ItemDescription
1The instrument adopts the measurement method of rough and precise measurement, to avoid the damage of mistake operation.
2Display the trend of resistor measurements.
3Automatically conduct comprehensive data analysis, and display the result value on the LCD screen.

4. Specifications
ResistorResolutionReference ResistorNominal RatioExcitation CurrentTest powerMeas. Accuracy (k=2 ,± μΩ/Ω)
1 mΩ10 pΩ1 Ω0.001:13.16 A10 mW5
10 mΩ100 pΩ1 Ω0.01:11 A10 mW3
100 mΩ1 nΩ1 Ω0.1:1316 mA10 mW2
1 Ω10 nΩ1 Ω1:1100 mA10 mW0.5
10 Ω0.1:1
10 Ω100 nΩ1 Ω10:131.6 mA10 mW0.5
10 Ω1:1
100 Ω1:10
100 Ω1 μΩ10 Ω10:110 mA10 mW0.5
100 Ω1:1
1 kΩ1:10
1 kΩ10 μΩ100 Ω10:13.16 mA10 mW0.5
1 kΩ1:1
10 kΩ1:10
10 kΩ100 μΩ1 kΩ10:11 mA10 mW1.5
10 kΩ1:1
100 kΩ1 mΩ10 kΩ10:1
    1. mA
1 mW2.5

5. General Specifications
Power supplyAC ( 220 ± 22 ) V,( 50 ± 2 ) Hz;
Warm up timeTwice the time since last warmed up, to a maximum of 60 minutes.
Maximum power consumption100 VA
Temperature performanceOperating temperature: 18 °C ~ 28 °C;
Storage temperature: -20 °C ~ 60 °C;
Humidity performanceOperating humidity: (20 % ~ 50%) R·H, non-condensing;
Storage humidity: (15 % ~ 80%) R·H, non-condensing;
WeightAbout 22.5 kg
Communication interfaceRS232,LAN
Dimensions485 mm(W)*355 mm(D)*220 mm(H) (Excluding handles and feets)

Company Profile
 