TM7900 Precision Volt-second Generator For Flux Meter Calibration

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 9000 ~ 25000

Description
 
TM7900 Precision Volt-second Generator For Flux Meter Calibration

1. Summary

TM7900 is a high-precision Calibrator for Flux Meter , which generates magnetic flux by a built-in high-precision Volt-second Generator (1Vs=1Wb). It calibrates flux-meters by comparing indicating value of the flux-meter with the value it generates. Compared with Mutual Inductance method, Volt-second method is featured with easier to use, better precision and traceability, it is applied for calibrate digital flux-meter and electrical integrator.
 

2. Features

  1. Accuracy Class: 0.005, 0.01, 0.02 or 0.05.
  2. Built-in high-precision voltage source and timer.
  3. Generate magnetic flux by integrating voltage and time. 
  4. Calculate and output voltage and time by setting magnetic flux value.
  5. Quick calibration in several seconds. 
  6. Unit switch: mWb, Wb, Vs, mVs, Mx.
  7. Support setting interior resistance and wire resistance.
  8. Eliminate interior resistance's influence by compensation.
  9. Large LCD screen and press-key.
 

3. Application

 Fast calibration of Digital Fluxmeter


 

4. Specifications

4.1 DCV Output

RangeAccuracy
±(ppm of reading + ppm of range) [1]
Class 0.05Class 0.02Class 0.01Class 0.005
2 mV200 + 200100 + 10050 + 10020 + 100
20 mV200 + 200100 + 5050 + 3020 + 10
200 mV200 + 200100 + 5050 + 3020 + 10
2V200 + 200100 + 5050 + 3020 + 10
Note [1] : (ppm = parts per million) (e.g., 10ppm = 0.001%).
  1. Regulating fineness: 0.005*Range, 0.001%*Range (class 0.005).
  2. Protective function: Short-circuit protection, Overload protection.
 

4.2 Adjustable Timer

Pulse outputAccuracy (±)
Class 0.05Class 0.05Class 0.05Class 0.05
0.1 s0.50.50.20.2
0.5 s0.50.50.20.2
1 s0.50.50.20.2
5 s0.50.50.20.2
  1. Pulse width output range: 0.1s ~ 5s
 

4.3 Standard Flux Output

Range
Accuracy
±(ppm of reading + ppm of range)
Class 0.05Class 0.05Class 0.05Class 0.05
Combination of voltage and time500 +0.1200 +0.1100 +0.150 +0.1
  1. Flux output range: 0.1mWb ~10Wb, 7-dights display.

5. General Specifications

Power supplyAC (220±22) V,(50±2) Hz
Temperature performanceOperating temperature: 0~50ºC;
Storage temperature: -20°C~70°C
Humidity performanceOperating humidity: 40%~80% R·H, no condensation
Storage humidity : < 95% R·H, no condensation
WeightAbout 4 kg
Communication interfaceRS232 interface
Dimensions180mm(width) × 270mm (depth) × 230mm (height)
 