TS2500 Magnetic Properties Measuring System for Electrical Steel Cores

Min.Order: 1
Product origin: Changsha, Hunan, China
Infringement complaint: complaintComplaint
US$ 25000 ~ 50000

Description
TS2500 Magnetic Properties Measuring
System for Electrical Steel Cores


*Above picture is only for reference, subject to actual delivery

1. Summary

TS2500 consists of excitation and measurement host, automatic measurement software. It's specially used to testing magnetic properties of motor laminated stator or electrical steel core.It can automatically measure the iron loss and other magnetic parameters of the motor stator core or other types of electrical steel core within a certain frequency range.
 

2. Features

  1. Electrical parameters calibration function.
  2. Testing frequency: 50Hz or 60Hz (400Hz,1kHz is optional).
  3. Testing mode: setting H to measure B or setting B to measure Ps.
  4. Testing in full frequency range with good accuracy and repeatability.
  5. Automatically testing with professional software.
  6. Automatically calculating magnetic parameters and curves.
  7. Complete curves drawing and data management functions.
  8. Modular design, easy to upgrade or maintenance.
  9. Multi-strand wire is customizable to simplify winding process.
 

3. Applications

Core Magnetic Properties Measurement
  1. Testing silicon core or stator core by direct winding.
  2. Setting multiple test points according to frequency and magnetic flux density B.
  3. Measurement Ps by locking B value.
  4. Magnetic parameters: magnetic field strength Hm, magnetic flux density Bm, specific total loss Ps, specific apparent power Ss, waveform coefficient FF, total loss P, etc.
  5. Draw magnetic characteristic curves: hysteresis loop, Ps-Hm,Ps-Bm curves, etc.
 

4. Specifications

4.1 Electrical Specifications

OutputMax. Power10 kVA (Optional)
Frequency50 Hz / 60 Hz…400 Hz…1 kHz (Optional)
ProtectOver current, Overload, Over temperature, Short circuit protection
MeasurementUncertainty of U and I (k=2)0.1%*RG [1]
Uncertainty of Power (k=2)0.3%*FS [2]
  1. Note: [1]RG is the range value, [2]FS = Voltage range value × Current range value.
  2. Note: The above indicators are typical cases, and the actual parameters should be planned in detail according to the samples tested by users.
 

4.2 Magnetic Specifications

Typical Frequency PointMagnetic ParametersTypical Test Point[5]Uncertainty (k=2) Repeatability
50Hz / 60HzPsP0.5 [2]... P1.5 2.0%0.5%
400 HzPsP0.5 [2]... P1.52.5%1.0%
1 kHzPsP0.5 [2]... P1.03.0%1.5%
  1. Note[3] : Test conditions : ( 23 ± 5 )ºC, demagnetize core before testing.
  2. Note[4] : Don't move the measured core and coils, repeated measurements for serval times, and calculate measurement repeatability.
  3. Note[5] : The test points are only reference. Actual performance of the measured core due to the material properties, weight, testing frequency, current, power load capacity.
  4. Note[6] : P1.0 refers to the specific total loss value when B=1.0T, and so on.
 

5. General Specifications

Power SupplyAC ( 380 ± 38 ) V,( 50 ± 2 ) Hz
Temperature PerformanceOperating Temperature:( 23 ± 5 )ºC;
Storage Temperature:-20°C~70°C
Humidity PerformanceOperating Storage:40%~80% R·H,non-condensing
Storage Humidity:< 80% R·H,non-condensing
 