Features
Model | Gentier 96E | Gentier 96R |
Thermal Block | ||
Sample capacity(wells) | 96 | |
Reaction Volume | 0-100 μL | |
Consumables | 0.2ml 96-well Plates(unskirt); 0.2ml 8-strip tubes ,0.2ml PCR single tube (Optical flat cap,clear,frosted,white tube) | |
Temperature range | 0°C-100°C | |
Heating/cooling method | Peltier | |
Max heating rate | 6.1°C/sec | |
Average heating rate | 4.5°C/sec | |
Max cooling rate | 5.0°C/sec | |
Average cooling rate | 2.8°C/sec | |
Temperature accuracy | ±0.1ºC | |
Temperature uniformity | ±0.1ºC | |
Gradient Range | 1°C-40°C | 1°C-40°C |
Gradient block | 12 row | 12 row |
Special temperature protocol | Gradient PCR, Long PCR, Touch Down PCR | Gradient PCR, Long PCR, Touch Down PCR |
Heat Lid | ||
Temperature range | Room Temperature-110°C | |
Optical System | ||
Excitation Source | 6 LEDs | 4LEDs |
Detector | Photodiode | |
Detection Position | Excitation and scan at top | |
Detection method | 6 channels scanning at the same time, no edge effect. | |
Detection time | 7 seconds for 96 wells for all channels | |
Range of excitation/emission wavelengths (nm) | 1.465 /510 (FAM,SYBR Green I,SYTO9,EvaGreen,LC Green) 2.527/ 563(HEX,VIC,TET,JOE) 3.580 /616 (ROX,Texas Red) 4.632/ 664( Cy5) 5.680 /730 (Alexa Fluor680) 6.465 /616 (FRET) | 1.465 /510 (FAM,SYBR Green I,SYTO9,EvaGreen,LC Green) 2.527/ 563 (HEX,VIC,TET,JOE) 3.580 /616 (ROX,Texas Red) 4.632/ 664( Cy5) |
Probe | Taqman Probe, Molecular beacons probe, Scorpion probe , FRET | Taqman Probe, Molecular beacons probe, Scorpion probe |
Multiplexing | Up to 6 targets | Up to 4 targets |
Fluorescence Linearity | r≥0.990 | |
Fluorescence Dynamic Range | Adjustable | |
Performance | ||
Sample Linearity | /r/≥0.999 | |
Sample repeatability | Ct value CV≤0.5% | |
Sample Dynamic Range | 1-1010 copies | |
Software Functions | ||
Data Analysis modes | Qualitative analysis,Absolute quantification, Relative quantification,Genotyping analysis,Endpoint analysis, Melt curve analysis, High Resolution Melting | |
Control modes | 1.10.4' touch screen 2. PC direct control 3. WLAN control(One PC can control max 10 units, and device can be controlled by any PC in the WLAN) | |
Sample Drawer | Touch screen commend | |
Data Stroage | Upload and download through USB disk,1000 results can be stored in machine | |
Power failure protection | Automatically start running experiments after power supply, no need wait PC software | |
Customize Report | Templates reserved ,report can customized | |
Administration Management | Administrator can set functions limits for users | |
Transport Locker | Automatically detect transport locker | |
Fault Management | Fault report and analysis, solution instruction | |
LIS connection | CSV,Excel,TXT format data output,open port for LIS connection | |
Others | ||
PC operating system | Win 7, Win 10 | |
Communication Port | 1 ethernet and 3 USB | |
Footprint (WxDxH) | 355mm X 480mm X 485mm | |
Weight | 30kg | |
Power usage | AC 100 to 125 V / 200 to 240 V (50/60 Hz). | |
power consumption | 900VA | |
Work environment | Temperature:10°C~30°C Humidity:20%~85% |