Name | Multi-parameter Patient Monitor | Model | NL070401001A-1 |
Place of Origin | CHINA | Certificate | CE ISO |
Standard Configuration | SpO2, PR/HR, NIBP, ECG, RESP | Optional Configuration | ETCO2, TEMP, Thermal printer |
Power Source | Electricity | Material | Plastic, ABS+PC Fireproof Materials |
After-sale Service | Online technical support | Color LCD Display | 7-inch color TET display,16:9 LCD display |
Power consumption | 100V-240V~,1.0-0.5A | Warranty | 1 Year |
Instrument Classification | Class II | Customization | Yes |
Application Range | Adult/Pediatric/Neonatal | Dimension and Weight | 175mm*140mm*300mm; 3.3kg |
Display | 7" real color TFT screen | Package | 360m*250m*300m; 4.5kg |
Operating environment | Displayed Parameters | ||
Temperature | 0~40ºC | Time | Battery-backed quartz crystal clock |
Humidity | 15%~80% | ECG | ECG Waveform Scale, displayed lead |
Pressure | 86.0kPa~106.0kPa | Heart Rate | Derived from ECG or SpO2 |
Power Requirements | NIBP | Pressure (systolic, mean and diastolic) | |
AC | 100~240V, 50Hz/60Hz | Pulse Oximeter | Pulse Rate, SpO2 waveform, and Oxygen saturation |
DC | Built-in rechargeable battery | Respiration Rate | Respiration rate derived from ECG |
Battery | 11.1V 24wh Li-ion battery | ||
2h operating after full charge | Trends | HR, RR, NIBP, TEMP and SpO2 | |
5min operating after low battery alarm | Temperatures | One channel | |
ECG | SpO2 | ||
5 Leads | RA, LA, LL, RL, V | Measurement range | 0~100% |
Lead mode | I, II,III, AVR, AVL, AVF, V | Resolution | 1% |
Gain | x1.25mm/mV, 2.5mm/mV, | Accuracy | 70%~100%±2% |
5.0mm/mV, 10mm/mV, | 40%~69%±3% | ||
20mm/mV (error≤5%) | |||
Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s , 50mm/s(error≤±10%) | Pulse Rate | |
Bandwidth | Surgery 1~25Hz | Measure/alarm range | 20~300bpm |
Monitor | 0.5~40Hz | Resolution | 1bpm |
Diagnostic | 0.05~130Hz | Accuracy | ±3bpm |
Sensitivity | >200μpVP-P | RESP | Oscillometry |
CMRR | ≥112dB | Method | Impedance between RA-ALL |
Differential Input Impedance | >5M | RESP impedance range | 0.3~3Ω |
ohm Electrode Offset Potential | ±300mV | Base Impedance range | 200Ω~4000Ω |
Polarization Voltage | ≥±500mV | Bandwidth | 0.3~2.5Hz |
Input Impedance | ≥5M | Gain | ×0.25, x 0.5, x 1, x 2, x 4 |
Calibration Voltage | 1mV±5% | RESP rate | |
ST Segment | Adult | 0~120BrPM | |
ST Segment Range | -2.0mV~+2.0mV | Neonatal/Pediatric | 0~150BrPM |
Accuracy | 0.02mV | Accuracy | 7~150BrPM(±2BrPM) |
0~6BrPM(unspecified) | |||
NIBP | Apnea Alarm | 10~40s | |
Method | Oscillometry | TEMP (optional) | |
Measure mode | Manual, Auto, STAT | Channel | 1 |
Measure interval in auto mode | 1~5/10/15/30/60/90/120/180/240/480min | Measure &alarm range | 0~50ºC |
Measure period in STAT mode | 5min | Resolution | 0.1ºC |
Pulse rate range | 40~240bpm | Accuracy(no sensor) | ±0.1ºC(0~50ºC) |
Measure & alarm range | |||
Adult: | SYS 40~280mmHg | Standard accessories | NIBP cuff & tube |
DIA 10~220mmHg | ECG cable & electrodes | ||
MEAN 20~240mmHg | SpO2 sensor | ||
Pediatric | SYS 40~220mmHg | Lithium-ion battery | |
DIA 10~160mmHg | Power cable | ||
MEAN 20~170mmHg | Operator's manual | ||
Neonatal | SYS 40~135mmHg | ||
DIA 10~100mmHg | Optional accessories | ||
MEAN 20~110mmHg | TEMP probe (skin) | ||
Static pressure measurement | Trolley bracket | ||
Range | 0~300mmHg | Hanging Bracket | |
Accuracy | ±3mmHg | Monitor recorder | |
Resolution | 1mmHg Accuracy | Vehicle-mounted power supply |