Model | BK-3100 | |||||
Throughput | 60 samples/hour | |||||
Principle | Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB | |||||
Channels | 2 | |||||
Parameters | 21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDW,PCT, P-LCR, P-LCC) | |||||
Histograms | 3 histograms (including WBC, RBC, PLT histograms) | |||||
Sample Volume | Whole blood: 10μl, Prediluted: 20μl | |||||
Calibration | Manual, auto calibration | |||||
Counting Modes | Venous, capilary, prediluted | |||||
Printing Model | Auto print, manual print | |||||
LIS | Support Bi-directional LIS | |||||
Language | Chinese, English, Spanish, French, Russian, Ukrainian | |||||
Reagent | Diluent 20L, Hemolytic agent 500ml, Probe cleanser: 50ml | |||||
Data Input | 9.7-inch touch screen, mouse, keyboard (optional) | |||||
Output | Internal printer, support external printer | |||||
Printer paper | 57*30mm | |||||
Interface | 4 USB ports, network port,RS232 serial port. | |||||
Storage | 100,000 results with histogram | |||||
Unclog | High voltage burn, reverse high-pressure flush | |||||
Work Temperature | 15-30°C | |||||
Power Supply | 100~240 AC, 50/60Hz, 150VA | |||||
Relative Humidity | 30~85 | |||||
Atmospheric Pressure | 70.0kPa-110.0kPa | |||||
Carry-over rate | Parameter | WBC | RBC | HGB | PLT | |
CV | ≤ 1.5% | ≤ 1% | ≤ 1% | ≤ 3% | ||
Background | Parameter | WBC | RBC | HGB | HCT | PLT |
Background | ≤ 0.3×109/L | ≤ 0.03×1012/L | ≤ 1 g/L | ≤ 0.5% | ≤ 10×109/L | |
Linearity | Parameter | Measurement range | CV | |||
WBC | (0.5~10.00×109)/L (10.10~99.9×109)/L | (±0.5×109)/L ±5% | ||||
RBC | (0.05~1.00×1012)/L (1.01~9.99×1012)/L | (±0.05×1012)/L ±5% | ||||
HGB | (2~70g)/L (71~300.0g)/L | (±2.0g)/L ±2% | ||||
PLT | (10~100×109)/L (101~999×109)/L | (±10.0×109)/L ±8% | ||||
Precision | Parameter | Whole blood (CV) | Pre-diluted (CV) | Measurement range | ||
WBC | ≤ 2.0% | ≤ 4.0% | (4.0-15.0×109)/L | |||
RBC | ≤ 1.6% | ≤ 3.0% | (3.5-6.0×1012)/L | |||
HGB | ≤ 1.6% | ≤ 3.0% | (110.0-180.0g)/L | |||
MCV | ≤1.0% | ≤2.0% | (70.0-120.0)fL | |||
PLT | ≤ 4.0% | ≤8.0% | (150.0-500.0×109)/L | |||
External Size(W*D*H) | 295*467*418mm | |||||
Packed Size(W*D*H) | 404*564*616mm | |||||
Net weight | 19 KG | |||||
Gross weight | 31KG |
1.Why choose BIOBASE?
We are the leading manufactuer in lab furniture and lab equipment, OEM & ODM service are offered. And we have our own overseas techinican and installation team.
2. How can we guarantee quality?
Always a pre-production sample before mass production;
Always final Inspection before shipment;
3. When can I get the price?
We usually quote as soon as we get your inquiry. If you need urgent reply, please tell us your email or Skype account, we will contact you ASAP.
4. What can you buy from us?
Laboratory Equipment (Biological Safety Cabinet,Fume Hood,Laminar Flow Cabinet,Autoclave,Freezer and Refrigerator)
5. Why should you buy from us not from other suppliers?
BIOBASE group, founded in 1999 and been through high and stable development for 15 years, who is now with revenues of one hundred million, 1200 employees in China and abroad offices, 22.4 acres land industry parks and 20000 square meters production base.