Medical Equipment Hea313 Infusion Syringe Pump Enteral Feeding Pump

Min.Order: 1
Product origin: Nanjing, Jiangsu, China
Infringement complaint: complaintComplaint
US$ 200 ~ 250

Description

Product Description

Medical Equipment HEA313 Enteral Feeding Pump


Technical specification

Infusion rate (ml/h)1~400 ml/hr in 1 ml increments
Mean flow velocity accuracy±10% of selected flow rate
Dosage limit(ml)0-9999 in 1 ml increment
Meanvolume accuracy±10%
Loading speed andmeanaccuracy (ml/h)1000±10%
Cleaning speed  andmeanspeedaccuracy(ml/h)1000±10%
Max volume flushandmeanspeed accuracy(ml)40±10%
Mx feeding pump pressure≥0.1 MPa
Occlusion alarm level (pressure)50KPa±10KPa
The longest occlusion triggering time of this device at minimum flow rate43min
The shortest occlusion triggering time of this device at intermediate flow rate85s
Medium speed running the pills from blocking the alarm threshold0.79ml
Recoverable Alarm1min50s-2min
Pause overtime alarm time1min50s-2min
Heating temperatureInfusion flow rate ≥20ml / h, 375px distance away from the heater outlet at 25-38 ºC
Use ofinfusionsetSee recommended infusion sets list
Air-in-line sensitivityDetestability of single size of air bubble ≥50uL
AlarmOcclusion alarm, Over heatedalarm, battery  depleted alarm
AlarmsAir in line,Prime, Flush, low power, failurecodes (ERxx), pause overtime, powerstate, infusion mode, work state, heaterstate.
NoiseMeasured at the highest flow rate on wooden table, the volume is lower than 50 dB at a distance of 1 meter
PowerAc: 100V-240V  50 Hz /60Hz
Built inbatteryA fully charged internal DC11.1V battery can continuously run the pump for approximately 5 hours on intermediate rate
Power consumption≤40VA
Power output≤20VA
The rated output voltage15V
Weight2.0kg
HeaterW-E01 ,cable length is 1.5m,Shield
Essential performanceThis product is suitable for patients with enteral nutrition infusion, the host cancontrol the infusion rate. With alarmfunction, with bubble detection piping and piping blockage detection. 
Classificationtype BF,class ΙΙ continuous volumetric feeding pump with internal battery; IPX1 (sealed drip-proof device)
Dimensions155mm(L)×170mm(W)×180mm(H)
Accessories1 user manual, 1 power cable, 1 heater (EA312/313 models only)

 

Company Profile

 

FAQ

1. What about the price level? 
We are confident that our prices are competitive compared with those at the same quality level, since we have stayed in this filed more than 20 years. And we always adhere to the principle of provide top quality products with competitive prices. We have already established our good fame in many countries, especially in Africa, Middle East, European, Asia and South American through year after year efforts. 

2. Why we find some prices are very high compared with some other supplier, and from the picture or it's model number, all looks the same?
As China is still a developing country, there are only a few products with patent rights, and copy products I spread very fast in the market. Of course, the prices of those copy products with cheap material and less quality control are relatively cheaper.

3. What is time of delivery? 
Since our products are produced according to each specific order, the delivery period will be 15 days after confirmation of order. If it is small or sample order, please do check with us about whether we have ready goods in stock to make delivery as soon as possible. 

4. What about quality assurance? 
We offer one year warranty period beginning from the time of shipment for medical equipment, Per minor problems, we can provide free spare parts for replacements, Per serious problems, we can make replacement of free charge. 

5. What is payment terms? 
T/T, Credit Card, L/C, Western Union, PayPal, MoneyGram. 

 

Scroll to Top