Area | Name | Model | Specification | Quantity |
Reagent preparation room | Laminar flow cabinet | BBS-DDC | 1040*615*1770mm | 1 |
Water purifier | SCSJ-20VF | 410*270*500mm | 1 | |
Refrigerator | MXC- 5V160 | 500*550*1300mm | 1 | |
Freezer | MDW-40V90 | 544*623*842mm | 1 | |
Mini centrifuge | Mini-7K | They are placed in the designated place for transportation and in the Laminar flow cabinet for use | 1 | |
Rotating mixer | MX-RL-PRO | 1 | ||
Pipette | MicroPette0.1-1000ul | 1 | ||
Balance | BA2204B | 1 | ||
Specimen preparation room | Biosafety cabinet | BSC-110IIB2-X | 1100*750*2250mmmm | 1 |
Refrigerator | MXC- 5V160 | 500*550*1300mm | 1 | |
Freezer | MDW-40V90 | 544*623*842mm | 1 | |
High speed centrifuge | BKC-TH16 | Equipped with 24*1.5/2ml angle rotor | 1 | |
Nucleic Acid Extraction System | BNP32 | 430*380*450mm | 1 | |
Water bath | SY-1L2H | 320*180*210mm | 1 | |
Rotating mixer | MX-RL-PRO | They are placed in the designated place for transportation and in the Laminar flow cabinet for use | 1 | |
Mini centrifuge | Mini-7K | |||
Pipette | MicroPette 0.1-1000ul | 1 | ||
Amplification& Analysis room | Fluorescence Quantitave PCR Detection System | FQD-96A (4-channel) | 410*386*352mm | 1 |
Washing room | Autoclave | MFS-B50II | 640*550*970mm | 1 |
1.Capsule |
*Single box structure , factory modular form a complete set *Dimensions: 12192mmx2438mmx2896mm (l/ D), standard container body, easy to transport |
2.Inside Cabin |
*The ceiling and internal partition walls are made of 50mm laboratory dedicated antibacterial purification board with a thickness of up to 0.426mm , which has good antibacterial effect, fire prevention, moisture-proof |
3.Area Arrangement |
*The PCR laboratory was designed in strict accordance with the three functional divisions: reagent preparation area, specimen preparation area and product amplification and analysis area, with independent buffer room.In addition, configure the washroom and equipment room. *The built-in PCR corridor facilitates the observation of the operation of each partition. |
4.Ventilation |
*Strictly follow the PCR laboratory air supply and exhaust design to ensure the air volume and ventilation frequency in the experimental area *The combined clean air conditioning unit adopts primary and medium efficiency two-stage filter air supply for the fresh air unit and high-efficiency filter for the terminal air supply port to carry out three-stage filter air supply to ensure the cleanliness of the room meets the requirements. |
5.Water Supply & Drainage System |
*The clean water tank and waste water treatment equipment are integrated between the equipment to maintain the normal operation of the laboratory under the condition of water cut o *Equipped with 260L capacity of clean water tank |
6.Laboratory Table |
*The mesa is 12.7 mm thick and the edges of the high-quality solid core physicochemical board are thickened to 25.4mm, which is resistant to corrosion , abrasion and strong acid and alkali. |
7.Controlling System |
*Visual intelligent control system , with room temperature and humidity, pressure gradient, lighting, uv lamp control |