Technical Parameter
Model | OK-MP10 | OK-MP20 | OK-MP40 | OK-MP60 |
Output | 10L/h | 20L/h | 40L/h | 60L/h |
Feed water requirements | Tap water,temperature 10~35ºC,pressure:0.15-0.35Mpa; TDS:≤200mg/L | |||
RO water quality |
| |||
Conductivity(25ºC) | 3-6μs/cm | |||
Iron rejection rate | 99%(new RO membrane) | |||
Organic rejection rate | >99%,when MW>200Balton | |||
Paeticles and bacteria rejection rate | >99% | |||
Ultrapure water quality |
| |||
Resistivity(25ºC) | 18.2MOhm-cm | |||
Bateria | <1CFU/ml | |||
Particle | >0.22um/ml<1um/ml | |||
AC requirement | 110/220V,50/60Hz | |||
Dimension(L*W*H-mm) | 410*300*500mm | |||
Weight(kgs) | 20kgs,30kgs,50kgs,70kgs | |||
Standard configuration | Main body,pressure tank |
The water purifying technology process is divided into source water pretreatment system,reverse osmosis system and ultra-pure system.
1. Source water pretreatment system: PP fiber filter + AC filter + Softener
2. Reverse osmosis system: It has high removal rate more than 99% for high valent ions, colloids, bacteria and organic matter (including endotoxin) with molecular weight greater than 300Dalton.
3. Ultra pure system: Ion exchange resin using H + exchange cation out the water in the filter ,using OH- exchange anions out the water , the replaced H + and OH- combined to produce H2O, ultra-purified column filling the United States Dow UPW-450 Ultra-pure water for nuclear grade resin, the water resistance of up to 18.2MΩ.cm.