Sy-G076-2 Digital Automatic Calibration PCA Ambulatory Infusion Pump

Min.Order: 1
Product origin: Guangzhou, Guangdong, China
Infringement complaint: complaintComplaint
US$ 88 ~ 250

Description
SY-G076-2 Digital Automatic Calibration Pca Ambulatory Infusion Pump


Specification details

1)Alarms:
· Occlusion, air-in-line, door open, end program, low battery, end battery,
· AC power off, motor malfunction, system malfunction, reminder alarm

2)Additional Features:
· Real-time infused volume / bolus rate / bolus volume / KVO rate,
· automatic power switching, mute key, purge, bolus, system memory,
· key locker, change flow rate without stopping the pump

 
Model
SY-G076-2
Pumping Mechanism
Curvilinear peristaltic
IV Set
Compatible with IV sets of any standard
Flow Rate
0.1-1500 ml/h (in 0.1 ml/h increments)
Purge, Bolus
100-1500 ml/h (in 1 ml/h increments)
Purge when pump stops, bolus when pump starts
Bolus volume
1-20 ml (in 1 ml increments)
Accuracy
±3%
VTBI
1-9999 ml
Infusion Mode
ml/h, drop/min, time-based
KVO Rate
0.1-5 ml/h (in 0.1 ml/h increments)
Alarms
Occlusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm
Air-in-line Detection
Ultrasonic detector
Wireless management
Optional
Additional Features
Real-time infused volume / bolus rate / bolus volume / KVO rate,
automatic power switching, mute key, purge, bolus, system memory,
key locker, change flow rate without stopping the pump
Occlusion Sensitivity
High, medium, low
Battery
9.6±1.6 V, rechargeable
Battery Life
5 hours at 30 ml/h
Power Supply, AC
110/230 V (optional), 50-60 Hz, 20 VA
Working Temperature
10-40ºC
Relative Humidity
30-75%
Atmospheric Pressure
700-1060 hpa
Size
174*126*215 mm
Weight
2.5 kg
Safety Classification
Class I, type CF

Scroll to Top