Inert Atmosphere Lab Glove Vacuum Box for Lithium Battery

Min.Order: 1
Product origin: Xiamen, Fujian, China
Infringement complaint: complaintComplaint
US$ 9800 ~ 12000

Description
Inert Atmosphere Lab Glove Vacuum Box For Lithium Battery
1.Model 
Super(1220/750/900)
Single station, single operation, integrated with a single column purification unit, PLC touch screen operation control, vacuum pump, bracket, a closed box, with operating surface tilt and detachable safety glass front window. Under normal conditions, namely the 20ºC constant temperature, a standard atmosphere, inert gas 99.999%, water and oxygen index less than 1 ppm.
 
2.Technology parameters                     
ItemParameter
Glove boxBoxBox size: (L*W*H)=1220×750×900mm
Box material: stainless steel 304, thickness 3mm
Inner surface: stainless steel wire drawing treatment
Outside surface: paint, white
Front windowTilt windows, transparent tempered safety glass, thickness 8 mm
Glove portsThe material is polyoxymethylene, O-ring seal, self-lubricating performance, easy to maintain, and good sealing performance
GlovesButyl rubber, thickness 0.4mm, diameter 8", length 32"
Filter:0.3 microns, a gas inlet and a gas outlet
ShelfStainless steel, Adjustable height
LightingLED lights, installed in front of each glass window
ConnectorStandby interface 3pcs, DN 40 KF, Power interface 1pc (220V)
Bigger transitive cabinChamberSize: diameter 360mm, length 600mm
Material: 304 stainless steel
Surface: inner surface wire drawing process, outer surface painting (white)
AccessarySliding tray: 304 stainless steel
Door: double door, anodized aluminum material, thickness 10mm, vertical operation, with lifting mechanism
Pressure gauge: Analog display
ControlSolenoid valve touch screen automatic operation
Smaller transitive cabinChamberSize: diameter 150mm, length 300mm (100mm length into the glove box)
Material: 304 stainless steel
Surface: inner: wire drawing process, outer: painting (white)
AccessaryStandard configuration: small cabin tray
Door: double door, thread lock
Pressure gauge: Analog display
ControlManual valve manual operation
Gas purification circulation systemPurification columnFunction: gas sealed, remove water, remove oxygen
Container material: 304 stainless steel
Purification materials: copper catalyst: 5kg, molecular sieve: 5 kg
Purification capacity: oxygen: 60L, water: 2Kg
Water and oxygen indicators: less than 1ppm
Circulatory systemWorking gas: nitrogen, argon, helium
Recycling capacity: integrated fan flow 90m³ / h, install frequency conversion
RegenerationOperation: PLC automatically controls the regeneration process
Regeneration gas: working gas mixed with hydrogen gas, (hydrogen 5-10%)
Vacuum pump8m³ / h, rotary vane pump, with oil mist filter, air vibration control
ValvesMain valve: DN 40 KF, electrically operated angle valve
Control valve: electromagnetic integrated valve
Control SystemFunctionIncluding self-diagnosis, power-off self-starting characteristics, with pressure control and adaptive function
Automatic control, cycle control, password protection
Pressure controlControl box, transition chamber pressure, cabinet working pressure +/- 10mbar can be set freely, beyond +/- 12mbar system automatic protection
PedalControl the box pressure, easy to operate boost and buck
Display systemUsing Siemens PLC touch screen, display running status, box pressure, system records etc
Vacuum system controlVacuum pump, can be manually or through the PLC to start, flow 8m³/h, vacuum of the transition chamber, and keep the box pressure balance, vacuum pump limit vacuum ≤ 2x10-1pa
Optional componentsWater analyzerMeasuring range: 0~500ppm platinum electrode structure (P2O5 structure, can reach the factory level by simple cleaning), a wide range of applications, especially for users such as lithium battery manufacturing and metal organic, can be restored to the initial state through the cleaning and regeneration process, avoiding one-time pollution and scrapping The problem.
Oxygen analyzerMeasuring range: 0 ~ 1000ppm
The use of ZrO2 sensor, to avoid the fuel cell life is short, and avoid the problem of it can not be exposed in the air
Organic solvent adsorberSize: diameter 237mm, height 407mm, thickness 0.3mm
Material: stainless steel 304,
Fill with 9kg of activated carbon. Connected with the main circulation pipeline, the circulating gas can first flow through the organic solvent adsorber device to absorb the organic solvent in the gas. Manual ball valve control can quickly change the adsorption material without destroying the circulating atmosphere.
Remove HF deviceDimensions: diameter 237mm, height 407mm, thickness 0.3mm
Material: 304 stainless steel
Fill with 9kg alumina particles. Connected to the main circulation pipeline, the circulating gas can first flow through the HF removal device to absorb HF in the gas. Manual ball valve control, can quickly change the adsorption material without destroying the circulating atmosphere
 

  
 