Hood Type HEPA Compact Air Filter

Min.Order: 1
Product origin: Liaocheng, Shandong, China
Infringement complaint: complaintComplaint
US$ 150 ~ 600

Description
SNYLI's Aluminium terminal HEPA box is a good choice for the air filtration system in clean rooms of class 1000.It adopts indoor replace design,the HEPA filter can be replaced inside the room.And the aluminum housing can help to make the installation work more easier.It can also be equipped with air diffuser.

SNYLI's Hood type HEPA filter is designed with replaceable structure as one type of HEPA filter terminal box. It's composed of outer housing, thermal insulation cotton, round or inlet duct, air diffuser plate, hanger, HEPA filter.
The entire housing of this Hood type HEPA filter is permanently fixed on the ceiling through the lifting lugs on the outer frame. When the replacement of HEPA filter inside is needed, It does not need to remove the connecting pipe, and the HEPA filter can be replaced directly from the room. It is simple, time-saving and labor-saving.The standard housing material is made of anodized aluminum profile,the standard height is only 180mm, which is convenient for handling and installation.This type of HEPA terminal box is suitable for various types of ceilings. The size of the air inlet can be selected.
 
Housing mateialAL/Glavanized steelFilter materialUltra fine glass fiber/PTFE
Separator:Hot melt adhesiveStd.EN1822:2009H13/H14
Max temperature:70ºCEfficiency:99.99% / 99.999% at 0.3μm
Maxrelative humidity:100%Inlet duct :Squre or circle or customized
 
Room side replaceable design
The internal HEPA filter and external housing are separated design. The HEPA filter can be replaced separately under the ceiling of the clean room instead of  be replaced in the interlayer of the clean room.

Compact design,Reliable sealing performance

It's a good choice when the height of the clean rooms is limited.

With excellet EPDM sealing strip,strict leakage test and filtration efficiency test help to garantee the product quatlity.



Products dimensions

Unit Model

Unit dimension

L×W×D(mm)

Filter Model

Filter dimension

 L×W×D(mm)

Rated air flow

     (m/h)

Initial P.D

(Pa)

FB-C10600180

1210×600×180

H4RHB70560070

1170×560×70

3000

249

FB-B70570180

1170×570×180

H4RHB30530070

1130×530×70

2750

FB-905600180

905×600×180

H4RH865560070

865×560×70

2200

FB-870570180

870×570×180

H4RH830530070

830×530×70

2000

FB-600600180

600×600×180

H4RH560560070

560×560×70

1400

FB-570570180

570×570×180

H4RH530530070

530×530×70

1250


Application
This Hood type HEPA filter is the ideal terminal filtration device for Class 1000,Class 10,000 air conditioning system, which can be widely used in the purification air conditioning system of electronics,medicine, health, chemical and other industries.

Insulated and non-insulated HEPA box housing
In the clean room, the supply air temperature is lower than the indoor temperature when in summer, while generally higher than the indoor temperature in winter. Therefore, it is necessary to insulate the housing of the ternimal HEPA filte box to ensure that the air supply parameters meet the design requirements. The surface of the outside housing is covered with PEF insulation cotton, which has good insulation performance.

Install caution
1) Before installation,HEPA terminal box's filtration efficiency ,air flow,pressure drop,leakage situation must be confirmed to meet the construction requirements of the clean room.

2)The HEPA filter box,the cleanroom inside and inside ceiling where the products will be installed should be cleaned completely.

3) The seam between the flange of the HEPA filter terminal box and the ceiling plate needs to be strictly sealed, especially where there are cracks, special treatment is required. The air inlet and air duct should be strictly connected and reinforced and sealed.

4)During installing, pay attention to the direction of the parts. The direction of the parts should not be wrong. For example, the arrow on the outer frame should be consistent with the airflow direction, and the crease direction of the filter paper should be perpendicular to the ground, etc., to avoid leakage, deformation and damage to the equipment.

5)After the installation is completed, the equipment should be checked. At this time, it depends on whether the HEPA terminal box can operate normally and whether there is any abnormality during operation. In addition, it is necessary to pay attention to the value of the equipment and check whether the values of various resistances and pressures are normal. Within the range, the installation of the HEPA filter box must pay attention to these details. As long as the details are well controlled and the installation of the high-efficiency air supply is done, the operation of the high-efficiency air supply will basically be smoother, and the use effect can also meet the satisfaction.
Product Tag: