Description

Aucma CFD-50 is the first WHO PQS prequalified (E003/079) MetaFridge product

MetaFridge Provides Reliable Cold Storage in Unreliable Power Conditions
  • Maintains vaccines at a stable 2-8°C temperature using patented ColdSafe Technology.
  • Allows 5 days of holdover at 43°C ambient temperature.
  • Can operate continuously on less than 2 hours of power per day at 43°C ambient temperature.
  • Requires less than 6 hours of power after installation to reach safe storage temperature with 2-3 days of holdover.
  • Grade A user-independent freeze protection.
Integrated Power Protection
  • Integrated protection during voltage spikes and brown-outs.
  • Internal voltage stabilizer allows compressor to use a wide range of voltage input.
Onboard Monitoring
  • Integrated 30 DTR display shows temperature and alarm history.
  • Holdover indicator allows users to proactively protect vaccines.
  • Visual and audible alarms for temperature excursions or improperly closed door.
  • Measures and records performance data to enable diagnosis of issues.
  • Optional telemetry module enables remote temperature monitoring and alerting.