Middle-End Anesthesia System Factory Wholesale With CE and ISO13485 Certification
Ventilation Model | |
PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, P-A/C, V-A/C, IPPV, PCV, PSV, SPONT/CPAP, SIGH, MANUAL | |
Ventilator parameter range | |
Tidal volume(Vt) | 0.20mL~2000 mL |
Frequence | 1 /min~100 /min |
Oxygen concentration | 21-100% |
I:E | 4:1~1:8 |
PEEP | 0 cmH2O~40 cmH2O |
Pressure triggering sensitivity (Ptr) | -20 cmH2O~0 cmH2O (Based on PEEP) |
Flow trigger sensitivity (Ftr) | 0.5 L/min~30 L/min |
SIGH | 0 (off) 1/100~5/100 |
Apnea ventilation | OFF,5s~60s |
Pressure limit | 20 cmH2O~100 cmH2O |
Monitoring parameter | |
Frequence(Freq) | 0 /min~100 /min |
Tidal volume(Vt) | 0~2500 mL |
MV | 0 L/min~99 L/min |
Oxygen concentration | 15 %~100 % |
Oscillogram | |
P-T (pressure - time) | |
F-T ( flow - time ) | |
P-V loop ( pressure - volume loop ) | |
V-T(Volume-time) | |
F-V loop(flow-volume loop) | |
F-P loop(flow-pressure loop) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Low voltage alarm for back up battery | <11.3±0.3V |
No tidal volume | ≤5mL within 6s |
High oxygen concentration alarm | 19%~100% |
Low oxygen concentration alarm | 18%~99% |
High airway pressure alarm | 20 cmH2O~100 cmH2O |
Low airway pressure alarm | 0 cmH2O~20 cmH2O |
High minute volume alarm | Adult (5 L/min~20 L/min) |
Low minute volume alarm | Paed (1L/min~15 L/min, 0~10 L/min) |
Continuous pressure alarm | (PEEP+1.5 kPa) over 16s |
Suffocation warning | 5s~60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
Working Condition | |
Gas source | O2, AIR |
Pressure | 280 kPa~600 kPa |
Voltage | 220V±22V |
Power frequency | 50Hz±1Hz |
Packing | |
Wooden case packing size(main engine) | L 560* W 560 * H 615 mm |
G.W. | 32KG |
N.W | 19KG |
Air compressor: |
• The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Department and other departments. |
• Professional design for adult, child and infant inhalation anesthesia and respiratory management with advanced ventilation modes. |
• Outstanding ergonomic design, it ranks high level in safty, stability and convenience as well as user experiences. |
• Simplicity: 4 static casters with self-locking function. |
• Precision in an anesthesia ventilator with multiple ventilation modes: IPPV, A/C, SIMV, SIGH, MANUAL |
• Flexible configurations able to customize your requirements. |
• International standard and advanced technology suitable for wide range use. |
• Compact interface and big screen give you better operating experience. |
• Over 5,000 units are installed in more than 200 countries. |
• Designed and manufactured by Superstar Medical with over 25 years of experience in ICU field. |
• 15" TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram. |
• High precision flowmeter, instantly know the fresh gas flow to your patient. |
• Integrated breathing circuit design, ensure easy operating and keep tidy. |
• Multiple working modes such as volume control and pressure limit, adapt to wide range patient |
• Vaporizer with temperature, pressure, flow compensation and self-lock function, keep safety anytime. |
• Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects |
• Pressure-time, low-time loop oscillogram and high precision ETCO2, O2 concentration show in real time. |
• Vital sign monitor and anesthetic gas monitor are optional. |
• ETCO2 module and Anesthesia Gas Scavenging System (AGSS) are optional. |
• 3 level alarm system, visual and sound alarm information |
• With multiple type of alarm, reminder and protection functions. |
• Advanced power management control technology |
• Built-in backup battery provide the emergency power supply to the unit. |
• Low O2 pressure alarm and N2O cut-off protection. |