Product Description

 

Single-chip microcomputer control, LCD Screen (128*64 Dots)

Large LCD screen, can display multiple sets of data.

Large memory space ,can store multiple sets of data and curve.

Auto zero,100% adjustment function

Auto wavelength adjustment

Auto filter changing, large sample pool (5mm ~ 100mm)

With up to ten standard sample for measurement function of standard curve establish

Through direct input K, B factor is established for quantitative measurement of standard curve

Can be directly input standard sample and the corresponding concentration values to establish the standard curve for quantitative measurement

Switch off power save standard curve parameter measuring set

Can be done any standard curve editor, convenient for use.

With general parallel printer interface or printer interface, can print title bar, measurement data, curve of standard sample curve, and the curve.

With USB interface

Can achieve more accurate and flexible measurement with PC control (optional) Technical Specifications

Product Parameters

 

Product description, technical parameters and configuration

Character 
Single-chip microcomputer control, Touch LCD Screen (7 inches)
Large LCD display can display multiple sets of data
Huge memory space to store multiple sets of data and curves
Auto zero,100% adjustment function
Auto wavelength adjustment
Auto filter changing, large sample pool (5mm ~ 100mm)
Standard samples having a maximum ten-point standard curve measurement functions built
Direct input by K, B factor for quantitative measurement standard curve
Can be directly input standard sample and the corresponding standard concentration values and establish a standard curve for quantitative measurement.
Switch off power save standard curve parameter measuring set
General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data, curve parameters, the standard curve and the curve sample points
With USB interface
PC control can be achieved through more accurate and flexible measurement requirements (standard) Technical Specifications


Technical Specifications

 
Model V-1500C
Wavelength Range190~1100nm
Optical circuit structureC-T type
Band Width2nm
Wavelength Accuracy±0.5nm
Wavelength Repeatability≤0.2nm
Photometric Accuracy±0.3% τ
Photometric Repeatability0.15% τ
Stray Light≤0.04% τ (340nm NaNO2)
Stability0.0005A/30min@500nm
Photometric ModeT,A,C,E
Wavelength SettingAutomatic
Photometric Display Range-0.3~3A
Display modeTouch LCD Screen (7 inches)
DetectorImport Silicon Photodiode
Light SourceImport Tungsten Lamp
Power RequirementAC 220V/50Hz 110V/60Hz
Power120W
Dimensions(W*D*H)480×350×220mm
Weight11Kg
 
 
 