Model# | DW- LC1620A | |
S pecifications | ||
DW- LC1620A Pump | Isocratic system | G radient system |
F low rate range | 0 .001-9 .999m L/min | |
F low rate accuracy | ≤±0 .5 %(1 m L/min , water, room temperature) | |
F low rate precision | RSD≤ 0 .08%(1 m L/min , water, room temperature) | |
P ressure range | ≤ 42MPa | |
DW- LC1620A UV Detector | S pecifications | |
Lamp | Cathodeon lamp | |
Wavelength range | 190-700nm | |
Baseline noise | ≤ 8×10-6AU | |
Baseline drift | ≤ 1 ×10-4AU/h | |
Detection L imit | 4×10-9 g/ml(Naphthalene/methyl alcohol solvent) | |
Wavelength accuracy | ± 1 nm |