Orthopedic Surgical Instrument Bone Ao Drill (BJ1102AO)

Min.Order: 1
Product origin: Shanghai, China
Infringement complaint: complaintComplaint
US$ -1

Description



Specification:
Output power105w
Speed(r.p.m) 0-1100
Oscillations (osc/min)/
Torque(Nm)2.9
Operation(V) voltage14.4
Sterilization135°C
Swings (°) /
Noise(db)≤40

Notes:
1.Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the 
   temperature is over 135 ° C ,the motor will be damaged.
2.Normal bone drill can only hold drill bits less than 6mm,acetabulum 
   reaming drill with big torque,rpm is around 300,canulate drill is used 
   with nail or K wire ,rpm is around 600.
3.The battery need to be get out of the pwer tool after surgery. Otherwise 
   the small automatic discharge of the electric will make the life of the 
   battery short.
4.The battery cannot be sterilized ,it is not autoclavable.
5.The cranial bur must be washed right after the surgery by purger in 
   order to aviod the blood stick in the bur,then we should make it dry.
6.The battery space in the handpiece of the tool should not have any 
   water,after sterilization.
7.The cranial bur blade is made of medical stainless steel,in order to 
   avoid broken,so it is flexible and not very hard,it is easy to be attrite, 
   we suggest each blade only use for 1 surgery.


Scroll to Top