Product Feature:
Model | LFZ-TSI-500 | LFZ-TSI-500R | LFZ-TSI-500A | LFZ-TSI-500AR |
Display | LCD | |||
Sharking frequency | 10-300rpm | |||
Sharking accuracy | ±1rpm | |||
Shake plate amplitude | Ф26mm | Ф35mm | ||
Temp. range | RT+5 ~65ºC | 4~65ºC | RT+5 ~65ºC | 4~65ºC |
Temp. adjust accuracy | ±0.1ºC | |||
Temperature uniformity | ±1 (at 37ºC) | |||
Temperature fluctuation | ±0.3ºC(at 37ºC) | |||
control method | P.I.D Microcomputer intelligent control | |||
Size / number of plate | 830×500mm/2pcs | |||
Maximum capacity of single layer | 100m×77or250ml×45or500ml×37or1000ml×18or2000ml×13 | |||
Single standard capacity | 100m×45or250ml×35or500ml×24or1000ml×12or2000ml×8 | |||
standard configuration | 2 layers(Lower fixing clamp 1000ml×12,Upper fixed clamp 500ml×24) | |||
Timing range | ||||
Power (kw) | 1.8 | 2.1 | 1.9 | 2.2 |
power supply | AC220±10% 50~60Hz | |||
Dimensions (W*D*H) | 1060*800*1470 | |||
Packing size (W*D*H) | 1190*895*1640 | |||
NW | 280 | 300 | 420 | 440 |
fittings | function |
True color touch screen program temperature control system | 1,Adopted 16 byte 4.3-inch true color touch screen computer control system, resistive touch screen, high touch accuracy, stability and reliability; 2,Control mode, setting mode and program mode selectable; Programmable program five kinds of modes, each mode can be set to 30 control steps, each step can be set from 0 to 999 hours 59 minutes; 3,You can draw real-time curve and the historical curve, and print screenshots; save up to 1 year historical data, and can be exported U disk backup or printout; choose to data or curve mode, real-time print measurement data; 4,Permission login system to ensure the operation of security, the timing of all control functions locked; 5,The standard configuration of illumination, and sterilization function; 6,optional buy our company matching computer software, real-time monitoring equipment through Ethernet |
Micro printer | Print control data according to the need |
485 interface remote control system | Can be based on the needs of the work of remote control and data analysis, export |
USB interface | According to the needs of the work, export of experimental data, does data analysis |