Description


 
Basic RequirementLaminar Air Flow should be Ceiling Mounted with 99.999% HEPA Filter (0.3 micron) efficiency according to GMP to avoid particulate & microbial contamination during filling process. It should comply DOP test and air velocity criteria .
DimensionsCustomized
TypeCeiling/Floor Mounted
MOCSS304
Noise Level60 to 65 db from 1 meter distance
Blower AssemblyStatically & Dynamically balanced with suitable rating & size
Air Velocity0.45 +/- 20% m/sec under HEPA Filter
DOP PortDOP/PAO port should be provided
ClassClass-100
PVC CurtainThree side pvc curtain,one side open, curtain length 1850 mm
down from the LAF
Air Flow PatternVertical down flow
DocumentationOperation & Maintenance manual should be provided. Drawing of the equipment showing front elevation, side elevation & plan view should be provided. IQ and OQ dully prepared and supplied.
 
 