Features:
1.Throughput: 60T/H
2.8 inch touch screen
3.20 parameters + 3 histograms
4.3 counting modes
5.100,000 sample results
6.Support LIS and external printer
Model | BK-6190 | ||
Throughput | 60Tests/hour | ||
Measuring Principle | Impedance for cell counting Cyanide-free method for HGB | ||
Aperture Diameter | WBC chamber:100μm,RBC/PLT chamber:70μm | ||
Parameters | WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR | ||
Histograms | 3 histograms(includingWBC, RBC, PLT histogram) | ||
Sample Volume | Normal:10μl; Predileted:20μl | ||
Calibration | Manual & auto calibration | ||
Counting Modes | Venous, capilary, prediluted | ||
Printer | Built-in thermal printer,USB to external printer(optional) | ||
Data Input | 8 inch touch screen, keyboard, mouse | ||
Output | RS232 port, USB port, VGA port. LIS system avaliable | ||
Storage | 100,000 results with histogram | ||
Carry-over rate | Parameter | CV | |
WBC | ≤ 2% | ||
RBC | ≤ 1% | ||
HGB | ≤ 2% | ||
PLT | ≤ 2% | ||
Background | Parameter | Background | |
WBC | ≤ 0.2×10^9/L | ||
RBC | ≤ 0.02×10^12/L | ||
HGB | ≤ 1 g/L | ||
HCT | ≤ 0.5% | ||
PLT | ≤ 10×10^9/L | ||
Precision | Parameter | CV | Measurement range |
WBC | ≤ 2.0% | (4.0-15.0×10^9)/L | |
RBC | ≤ 1.9% | (3.5-6.0×10^12)/L | |
HGB | ≤ 1.9% | (110.0-180.0g)/L | |
MCV | ≤ 0.4% | (80.0-110.0)fL | |
PLT | ≤ 4.0% | (100.0-500.0×10^9)/L | |
Linearity | Parameter | Measurement range | CV |
WBC | (0-5.9×10^9)/L (6.0-99.9×10^9)/L | (±0.3×10^9)/L ±5% | |
RBC | (0-0.99×10^12)/L (1.0-9.99×10^12)/L | (±0.05×10^12)/L ±5% | |
HGB | (0-99.0g)/L (100.0-300.0g)/L | (±2.0g)/L ±2% | |
PLT | (0-99×10^9)/L 100-999×10^9)/L | (±8.0×10^9)/L ±10% | |
Power supply | AC 220±10%,60/50Hz, 110±10%,60Hz | ||
Package Size | 580*470*590mm (W*D*H) | ReagentPackage Size | 450*410*310mm (W*D*H) |
Gross Weight | 27kg | ReagentGross Weight | 23kg |