Rotary Evaporator with Water / Oil Bath

Min.Order: 1
Product origin: Zhengzhou, Henan, China
Infringement complaint: complaintComplaint
US$ 3 ~ 5380

Description

Product Description


Technical Parameter
MODELR-1005R-1010R-1020R-1050
Optimum ambient temperature (ºC)5~25
Power (V/Hz)Single-phase 110/60Single-phase 220V/60HZ
Speed regulation of main engineDigital variable-frequency stepless speed regulationDigital direct current stepless speed regulation
Rotate speed of main engine (rpm)10-14020-13020-110
Rotary motor4060120
Temperature control (ºC)Digital temperature control, constant temperature -99ºC
Maximum vacuum degree (Pa)399.9Pa(under 3mmHg)
Rotary flask (L)5 Flange mouthØ5010 Flange mouthØ9520 Flange mouthØ12550 Flange mouth Ø125
Receiving flask (L)351020
Condenser pipeVertical dual coil pipeUpright, main condenser + auxiliary condenser, efficient tri-backflow condenser pipe
Evaporative capacity (L/h)Water ≥2 Alcohol ≥4Water ≥3.2 Alcohol ≥6.5Water ≥5 Alcohol ≥11Water ≥9 Alcohol ≥19
Water bath pot materialStainless steel
Water bath pot size (mm)Φ280×170Φ350×220Φ450×250Φ550×320
Lifting functionElectric liftingElectric + manual lifting
Lifting stroke (mm)0-1500-1600-1900-180
Complete machine power (kW)2.24.86.38.3
Dimension (mm)765×400×1070920×550×17001250×600×21001320×770×2340

Features of 50L Rotary Evaporator:

1.Rotating body is composed of a reliable three-phase asynchronous motor and a transducer control device; 
2.Double sealed by PTFE and fluorine rubber sealing ring to keep high vacuum degree, separate gas and liquid and maintain no backflow;
3.Adopt main and vice efficient condenser, there has automatic switching device between vice condenser and receiving bottle. Below receiving bottle, equipping with tapping valve, which makes it convenient to recycle solvent and the recovery rate approach to 100%; 
4.Rotary bottle is connected by nut, that is easy to install and dismount; 
5.Continuously charging and recycling material; 
6.Vacuum system is equipped with vacuum meter, choose best working vacuum degree to low boiling point materials; 
7.Heating kettle is applicable for both water and oil, digital thermostat control and manual up and down, closed resistance wire heater, temperature control by K type transducer. Besides, heating groove can move left and right; 
8.Has the function of dry burning resistant protection and safe temperature loop; 
9.Smooth starting, can positive and negative inversion operate; 
10.Durable design, can choose plate bursting disk glass component as needed.

Corollary Equipment

1. Rotary vacuum evaporator matching use with circulating water vacuum pump to offer a vacuum environment for reduced pressure distillation, they are widely used for vacuum filtration, vacuum evaporation, vacuum degassing, etc..
2. Rotary vacuum evaporator matching use with GDSZ series high and low temperature circulating pump to quickly cooling or warming, the circulating pump is an efficient equipment for quick temperature adjustment. 
3. Rotary vacuum evaporator matching use with DLSB series low temperature cooling liquid circulating pump to conduct chemical reaction and medicine storage which requires low temperature environment.

Related Products




Workshop





 

Q: What's the difference between vertical and inclined condenser?

A: They have no difference essentially. Vertical condenser is gradually popular because of its small space usage and convenience for transportation and use. Most high-capacity instruments use vertical condenser.

Q: What factor decides the working efficiency of distillation rotary evaporator?

A: For same type equipment, mainly includes: steam temperature, vacuum degree and cooling water temperature. Restricted by some material's thermolability, the steam temperature may have certain limitation when working, so high vacuum degree and low cooling water temperature are two main factors for efficiency improving.

Q: How to select all kinds of rotary evaporator?

A: As for structure, our R series products all have excellent corrosion resistance ability and leakproofness ability. According to the solvent volume, 2L, 3L, 5L's machine is fit for lab and sample test; 5L, 10L, 20L's machine is fit for medium experiment; 20L, 50L's machine is fit for medium experiment and production, especially suitable for the extraction of metal ionavoiding polluted material. 