Description
LCD display
Touch screen is optional
Wavelength Range: 340nm~700nm,
standard:340,405,492,510,546,578,630nm,1 free position for optional.
Wavelength Accuracy:±2nm
Input:keyboard input or touch screen
Measurement time: 0~999s programmable
Delay time:0-999s programmable
Spectrum bandwidth: 6-12nm
Cross-contamination:≤1%
Storage:up to 1000,000 test results
Sampling: 500-9999μl
Absorbancy: 0.3-3.0ABS
Lineary: less than 2% within 2ABS
Absorbancy repeatability: CV≤1%
Colorimetric Pool :Titanium alloy
Reagent: Open system,400μl~800μl.
Microflowcell volume: 32μl
Stability: 20 mins for warm-up,then≤2m ABS/H
Measure Repeatability:≤2mABS
Test Method: End-point, Kinetic, Two-point, Standard Curve and Single
Double Wavelength: Exam, absorbance
GW: 10KG
Wavelength Range | 340nm~700nm,standard 340,405,492,510,546,578,630nm, 1 free position for optional. |
Wavelength Accuracy | ±2nm |
Input | keyboard input or touch screen |
Measurement time | 0~999s programmable |
Delay time | 0-999s programmable |
Spectrum bandwidth | 6-12nm |
Cross-contamination | ≤1% |
Storage | up to 1000,000 test results |
Absorbancy | 0.3-3.0ABS |
Sampling | 500-9999μl |
Lineary | less than 2% within 2ABS |
Absorbancy repeatability | CV≤1% |
Colorimetric Pool | Titanium alloy |
Reagent | Open system,400μl~800μl. |
Microflowcell volume | 32μl |
Stability | 20 mins for warm-up,then≤2m ABS/H |
Measure Repeatability | ≤2mABS |
Test Method | End-point, Kinetic, Two-point, Standard Curve and Single |
Double Wavelength | Exam, absorbance |