Biobase China High quality/High cost performance Bk-UV1800 Automatic Single Beam UV/Vis Spectrophotometer with Silicon Photodiode Detector for Lab

Min.Order: 1
Product origin: Jinan, Shandong, China
Infringement complaint: complaintComplaint
US$ 801 ~ 1500

Description
Product Description

1.Features

 * Features:

*With GLP self-check function, check the wavelength accuracy and Photometry accuracy

*Wavelength calibration, wavelength setting, change lamp source and dark current calibration automatically

 

*With Si02 coating optical mirror, reducing the pollution from outside fully

*PC Software to expand the applications to Quantitative, Multi-Wavelength and Kinetics, Spectrum Scanning, *DNA/Protein test for PC Series

* Basic Functions:

1, Photometry

    Test Abs., Transmittance and Energy by fixed wavelength

2, Quantitative

     Linear fit and Linear fit through zero two modes

     a. Coefficient, Standard Sample input and Standard Sample read three modes to establish standard curve            

     b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings

     c. Can save 240 group curves, can test 240 data in each curve

     d. Double wavelength, Triple wavelength test functions

3, Kinetics

    Used for time course scanning or reaction ratecalculations A/t,can search all data.  

4, Multi-Wavelength

    Can test Transmittance and Abs. with 8 different wavelengths at most 

Product Parameters

2.Technical Parameters

Model

BK-UV1800

BK-UV1800PC

Optical System

Single Beam, Grating 1200 lines/mm

Wavelength Range

190~1100nm

Spectral Bandwidth

2nm

Wavelength Accuracy

± 0.5nm

Wavelength Repeatability

≤ 0.3nm

Photometric Accuracy

± 0.002A (0~0.5Abs), ± 0.004A (0.5~1.0Abs), ± 0.5% T (0~100% T)

Photometric Repeatability

0.001Abs (0~0.5Abs), 0.002Abs (0.5~1.0Abs), ≤ 0.2% T (0~100% T)

Stray Light

≤ 0.05% T @ 360nm; 220nm

Stability

± 0.001A / h @ 500nm

Baseline Flatness

± 0.002A

Noise

± 0.0005A

Display

480 * 272 65 thousand true color TFT LCD

Photometric Mode

T, A, C, E

Photometric Range

0~200% T, -0.301~3.0A

Detector

Silicon Photodiode

Light Source

 Lamp

Input

Membrane Keypad

Output

USB  Print and data output, Connect PC

Standard Accessories

10mm glass cuvette *4 units      10mm quartz cuvette *2 units

Optional Accessories

Single cell holder,General cell holder,Test tube holder,Solid sample holder lamp,Halogen lamp,4-place Cell holder,Printer,Water-jacketed cell holder,Reflection accessory 5°,Auto 8-cell changer,Micro cell holder,Water-jacketed cell holder,Constant-temperature system

 

Power Supply

AC110V/220±10% ,60/50Hz

Packing Size(mm)

620*500*370

Gross Weight(kg)

18

Company Profile

 

Packaging & Shipping

 

Contact me

 



Product Tag:
Related categories:
Scroll to Top