Basic range(mH2O) | 3.5 | 7 | 10 | 20 | 35 | 70 | 100 | 200 |
Overpressure(mH2O) | 5 | 10 | 15 | 30 | 50 | 100 | 150 | 300 |
Note: The min.shrinked available range is 1/5 of basic range; For the other pressure range between min.range and max.range, but not indicated on the above table, we enlarge the above basic pressure range to produce (overpressure unavailable); |
Total accuracy | ±0.075%FS(min.) ±0.1%FS(typ.) ±0.25%FS(max.) |
Stability | ±0.2%FS/year |
Compensation temp. | (-10~70)ºC |
Operation temp. | (-10~80)ºC |
Storage temp. | (-40~100)ºC |
Power supply | (10~28)VDC(RS485), (12~30)VDC(HARTâ) |
(10~12)VDC(Intrinsic safe version) | |
Output | (4~20)mADC |
·RS485 communication interface(Custom or MODBUS protocol) or HARTâcommunication protocol | |
Load | (4~20)mADC output,(U-12V)/0.02A(Ω) |
RS485 bus line could connect with 99 transmitter | |
Insulation | 100MΩ/50V |
Shock | 20g,(20~5000)Hz |
Impact | 20g,11ms |
Protection | IP68(Transmitter part);IP65(Junction box part) |
Media compatibility | Diaphragm: stainless steel 316L |
Housing: stainless steel 1Cr18Ni9Ti | |
O-ring: Viton | |
Weight | ~160g |
MPM4700 | HARTâ Intelligent Level Transmitter | ||||||||||||
Range | Pressure range :0~3.5200mH2O | ||||||||||||
[0~XmH2O]L | X=actual pressure range, L=cable length , suggested L-X=(1~2)m | ||||||||||||
Code | Output signal | ||||||||||||
E | 4~20mADC | ||||||||||||
R4 | RS485 communication interface, Custom protocol | ||||||||||||
R8 | RS485 communication interface, MODBUS communication protocol | ||||||||||||
H | HARTâ communication protocol | ||||||||||||
Code | Construction material | ||||||||||||
Diaphragm | Interface | Housing | |||||||||||
22 | SS 316L | SS | SS | ||||||||||
24 | SS 316L | SS 316L | SS 316L | ||||||||||
Code | Others | ||||||||||||
C1 | M20×1.5 male, face type seal | ||||||||||||
C5 | M20×1.5 male, waterline seal | ||||||||||||
Yb | Aluminum connection box without display | ||||||||||||
Yc | MS200 water-proof connection box(default) | ||||||||||||
Yd | PD140 lightening-proof protection device | ||||||||||||
F1 | Fixed flange | ||||||||||||
i | Intrinsic safe version ExiaIIBT6 Ga | ||||||||||||
T | Ship-use | ||||||||||||