Best Selling Automatic Glycated Hemoglobin Hba1c Analyzer

Min.Order: 1
Product origin: Anqing, Anhui, China
Infringement complaint: complaintComplaint
US$ -1

Description
Specifications
Basic Parameters
Testing Method
Low pressure ion exchange liquid chromatography
Testing Item
Glycated hemoglobin HbA1c
Testing Scope
3% ~ 18%
Testing Parameters
Precision (CV) ≤3%, Accuracy (V) ≤1.50%
Testing Time
4 minutes (reporting), Full-processing 6 minutes (including cleaning column and recovery time)
Sample type
Venous blood (EDTA anticoagulation),Peripheral finger blood
Sample volume
5μl whole blood for each sample
Functional Parameters
Photometer
415nm LED Integral flow colorimeter
Sampling mode
Full-auto 25-position sample turntable (20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position)
Testing mode
Auto test for batch, single test for emergency
Calibration
Optional selection of 2 points or 4 points calibration, optional selection of manual or automation, Equipped with 2 groups of high and low level calibrators
Thermostatic control
Column and reagent thermostatic control: 25 ± 0.5ºC
Reagent matching
One piece of column and reagent consumable/300 pieces of testing samples
Reagent allocation
Each standard package includes A, B, C, D eluent, calibrator, hemolytic agent, pump tubing
Output Parameters
Display
320x240 LCD Graphic Display, displaying real-time testing curve
Printer
58mm Thermal printer, printing testing curve and report
Report Output
IFCC concentration value, NGSP area percentage, ADAG average glucose
Data storage
1000 test report (including testing curve)
Communication interface
RS232 communication interface, connecting to HIS/LIS system
Working Parameters
Power Supply
AC220V±22V 50Hz 65VA
Size
408mm×390mm×230mm
Weight
8Kg
Working Environment
Temperature 12ºC-30ºC,Relative humidity ≤85%

Features
* Accurate Principle of Methodology
Adopts classical and accurate principle of methodology - Ion exchange liquid chromatography, It is the gold standard of HbA1C analysis, and it is the only analysis method to really separate HbA1C directly by measuring the piecemeal absorbance through continuous test on line,? and obtain the correct area percentage with integration.
* Accurate separation 4-gradient elution
The novel 4-gradient elution for HbA1C can separate accurately glycated hemoglobin with 4-gradient elution of corresponding concentration reagent aiming at HbA1C instead of routine ?elution process produced by high and low concentration mixture.
* High Separation Liquid Chromatographic Column
High separation liquid chromatographic column made of imported resin with volume of Φ 9mm x 45mm and weight of 2.5g which is 15 - 20 times greater than general micro chromatographic column. High efficiency chromatographic column for 300 tests ensure the accuracy of test results.
* High Sensitivity 415nm LED Integral Photometer
High sensitivity 415nm LED integral photometer has the characteristics of correct wavelength, stable light source, full aluminum alloy structure, high anti-interference performance, multi-lens focusing, micro cuvettes and high sensitivity. It can record accurately analysis curve.
* Supply with Original Calibrator
The international standard value tracing is adopted. The authoritative reference material for quality control is used to transfer the numeric value. Each set of reagent is supplied with two sets of calibrator for proper calibration so as to ensure that the test results are correct and reliable and thoroughly to avoid individual error caused by factor calibration.
* Precision Chromatographic Column and Thermostatic Apparatus
Precision chromatographic column and reagent thermostatic control apparatus ensure that Chromatographic column and reagent are not affected by environment temperature and effectiv