Pet Hospital Devices Veterinary Anesthesia Ventilator with Touch Screen

Min.Order: 1
Product origin: Shanghai, China
Infringement complaint: complaintComplaint
US$ 1000 ~ 2000

Description

Pet Hospital Devices Veterinary Anesthesia Ventilator with Touch Screen

Basic description

Parameter

Description

Display

8 " TFT, with touch screen.

Drive gas

Gas type

O2 or Air

Inlet pressure

280-600 kPa.

Max flow

≤ 120 L / min.

Operating Mode

VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT.

Optional: SIMV-PRVC, PRVC

Waveform

Waveform: 

Pressure,Flow-rate,Volume,CO2(optional),Pleth(optional)

Loop:PV loop, PF loop,FV loop

Safety Pressure

System pressure does not exceed 10 KPa.

Battery Type

Li-ion,work time more than 3 hours

 

Parameters setting range

Parameter

description

Tidal volume

Range: 50 - 1500ml; optional 10 - 1500ml

Increment: 10 ~ 100 mL: 5mL;

100 ~ 1500 mL: 10mL;

Respiratory rate

Range: 1 ~ 100 bpm; increment: 1 bpm.

Inspiratory time

Range: 0.1 ~ 10.0 s; increments: 0.1 s.

Respiratory ratio

Range: 4:1 to 1:10; increments: 0.5.

Percentage of inspiratory pause

Range: 0 to 50%; Increment: 5%

PEEP

Range: OFF, 3 ~ 20 cmH2O; Increment: 1 cmH2O.

Pressure Support

Range: 0 ~ 70 cmH2O; Increment: 1 cmH2O.

Pressure Control

Range: 5 ~ 70 cmH2O; Increment: 1 cmH2O.

Flow trigger

Range: 0.5 ~ 20L/min; increments: 0.1L/min.

Pressure Trigger

Range: 0 ~ 20 cmH2O; increments: 0.1 cmH2O.

PSV Insp Termination Level

Range: 5 ~ 80%; increments: 5%.

 

monitored parameters

Parameter

Description

Inspiratory tidal volume

Range: 0 ~ 2500 mL; Resolution: 1 mL. Error of ± 20mL or actual value ± 15%, which ever is greater

Expiratory tidal volume

Range: 0 ~ 2500 mL; Resolution: 1 mL. Error of ± 20mL or actual value ± 15%, which ever is greater

Minute ventilation

Range: 0 ~ 60 L / min; Resolution: 0.1 L / min. Error of ± 1L/min or actual value ± 15%, which ever is greater

Spontaneous minute ventilation

Range: 0 ~ 60 L / min; Resolution: 0.1 L / min. Error of ± 1L/min or actual value ± 15%, which ever is greater

Respiratory rate

Range: 0 ~ 100 bpm; Resolution: 1 bpm. Error of ± 2 beats / min or actual value ± 10%, which ever is greater

Spontaneous breathing frequency

Range: 0 ~ 100 bpm; Resolution: 1 bpm. Error of ± 2 beats / min or actual value ± 10%, which ever is greater

Respiratory ratio

Range: 30:1 to 1:150; resolution: 0.1. Error of ± 20%,

Peak airway pressure

Range: 0 ~ 100 cmH2O; Resolution: 1 cmH2O. Error of ± (2% + 4% of full scale actual reading)

Mean airway pressure

Range: 0 ~ 100 cmH2O; Resolution: 1 cmH2O. Error of ± (2% + 4% of full scale actual reading)

PEEP

Range: 0 ~ 100 cmH2O; Resolution: 1 cmH2O. Error of ± (2% + 4% of full scale actual reading)

Inspiratory plateau pressure

Range: 0 ~ 100 cmH2O; Resolution: 1 cmH2O. Error of ± (2% + 4% of full scale actual reading)

FiO2(optional)

Range: 15 to 100%; Resolution: 1%. Error is ± (2.5% +2.5% of full scale actual reading)

Compliance

Range: 0 ~ 300 mL/cmH2O; resolution: 1 mL/cmH2O. Error of ± 20% or ± 5 mL/cmH2O, which ever is greater

Airway resistance

Range: 0 ~ 600 cmH2O / (L / S); Resolution: 1 cmH2O / (L / S). Error of ± 20% or ± 5 cmH2O, which ever is greater

EtCO2 (optional)

Range: 0 ~ 100 mmHg; Resolution: 1 mmHg. Error is ± (0.43% of the volume percentage + 8% of the gas concentration) is equivalent to the range of optional units to monitor KPa and mmHg.

Inhalation of carbon dioxide (optional)

Range: 0 ~ 100mmHg; Resolution: 1 mmHg. Error is ± (0.43% of the volume percentage + 8% of the gas concentration) is equivalent to the range of optional units to monitor KPa and mmHg.

MAC values(optional)

Range: 0 to 10; resolution: 0.01