Model | KS-RS606 |
Analytical principle | Energy Dispersive X-Ray Fluorescence Analysis |
Analytical elements | Elemental analysis range Sulphur (S) - Uranium (U) any element. |
Can analyse copper alloys, iron alloys, zinc alloys, tin-lead alloys, precious metal analysis, etc. | |
Quantitative testing for lead (Pb), mercury (Hg), cadmium (Cd), chromium (Cr), total bromine in polybrominated biphenyls (PBB) and polybrominated diphenyl ethers (PBDE), and bromine (Br) and chlorine (Cl) in halogens in ROHS and halogens. | |
Up to 5 layers of plating can be analysed and up to 24 elements can be analysed at one time. | |
Working conditions | Operating temperature: 18-22°C |
Relative humidity: 40% - 60% | |
Power supply: AC 220 V ± 10 %, 50/60 Hz | |
Measurement conditions: Atmospheric environment | |
Lower limit of detection | Pb/Cd/Cr/Hg/Br/Cl<=2ppm |
Test sample time | 50-300 seconds adjustable |
Test content range | 1ppm - 99.99% |
Wide dynamic range of analytical measurements from 0.005μm to 50μm | |
Sample shape | Any size, any irregular shape |
Sample type | Plastics / Metals / Films / Powders / Liquids |
X-ray tube | Tube Voltage:0~50 kV |
Tube current: max. 1 to 1000 μA max. power: 50 W | |
Filament voltage: 2.0V | |
Filament current: 1.7 A | |
Ray take out angle: 12° | |
Target material: tungsten target | |
Beryllium window thickness: 400um | |
Over-temperature and overload protection, automatically enters standby mode when not working to extend service life | |
Illumination diameter | Collimator 8mm\2mm\1mm \0.5mm |
Detector | Amptek Si-PIN high sensitivity detector, USA |
Best resolution: 149±5eV. | |
Spellman High Voltage Power Supplies, USA | Input Voltage: DC +24V±10% |
Input current: 4.25A (max) | |
Output voltage: 0 -50KV & 1mA | |
Maximum power: 50W | |
Voltage regulation: 0.01% (from no-load to full-load) | |
Current regulation rate: 0.01% (from no load to full load) | |
Ripple voltage: 0.25% of the maximum output voltage at the peak ripple voltage under rated output voltage conditions. | |
8 hour stability: ≤ 0.05%. | |
Heat dissipation system | There are several heat sinks at the front and rear of the body to dissipate heat, three fans inside the body for air cooling, an X-ray tube for coating high temperature heat reduction resin after the heatsink is wrapped, and then one fan to blow air on it to dissipate heat. |
Radiation protection systems | A combination of soft and hard radiation protection design. Triple treatment: special radiation treatment of the X-ray tube, double lead skin wrapping of the sample cover, radiation warning of the software (the instrument will automatically shut down all power when the sample cavity is opened by mistake by personnel during work to ensure the safety of the operator), obtaining the radiation safety licence stipulated by the state, the radiation dose is lower than the national standard. |
Filters | Eight new filters with automatic selection settings |
Sample monitoring systems | The 30x5 megapixel colour CCD camera allows real-time observation of the sample status in the software test interface and micro-motion adjustment of the sample position prior to placing the sample for testing, and the pattern is clearly presented in the test report after the sample has been tested. |
Software systems | Quantitative analysis and calculation, using theoretical Alpha coefficient method (NBS-GSC method), integrated FP, method empirical coefficient method, for the accurate testing of RoHS halogen-free elements (Pb, Cd, Hg, Br, Cr, Cl). Multi Easy-to-use, one-touch interface, generates customised reports and automatically outputs PDF, EXCEL and other report formats. |
Rated power | 0.35kW |
Weight | 50Kg |
Dimensions | 650(W)×450(D)×450(H)mm |
Sample chamber size | 350(W) x 300(D) x 150(H)mm Large sample lid for open lid open testing. |
Standard configuration | One set of Amptek Si-PIN detectors |
One x-ray tube with tungsten target with special radiation protection treatment and side window | |
One Spellman high voltage power supply from USA | |
Fully enclosed lead plate double sample chamber | |
One high precision CCD camera | |
A set of software for Alpha coefficient method (NBS-GSC method), integrated FP, method empirical coefficient method | |
One sample box, one sample cup for testing, one roll of plastic film for testing | |
One EU PE# specimen and one silver calibration specimen | |
Operating system | WINDOWS7 |
Internal memory | 4G |
Display | 19" SVGA(LCD) |
Hard disk | 1T |
Printer | HP colour inkjet printer |
CPU | Intel i3 |