Hach Surface Scatter® 7 Sc (SS7 sc) Sensor Assembly (Instrument Only) Turbidimeter

Min.Order: 1
Product origin: Shenzhen, Guangdong, China
Infringement complaint: complaintComplaint
US$ 4029

Description

Hach Surface Scatter® 7 sc (SS7 sc) Sensor Assembly (Instrument Only)

Outstanding results in high temperatures: SURFACE SCATTER 7 sc

The Surface Scatter 7 sc High Range Turbidimeter with sc200 controller offers superior performance features. It measures a turbidity range from 0 to 9999 NTU. Includes calibration cylinder, 4000 NTU Formazin, installation accessories, and instruction manual. Note: only one SS7 sc sensor can attach to one sc200 controller, if a second different sc sensor is required for this configuration please contact the Hach Company - Process Technical Group for further details.
 

Benefits

Virtually maintenance-free, optics never touch the sample

The Hach Surface Scatter 7 sc Turbidimeter (SS7) is uniquely designed so that the light source and photocell never come in contact with the sample. In fluids with high loads of suspended solids this makes sample cell cleaning and replacement unnecessary.

Durable-manufactured with corrosion/heat-resistant materials for extended life

All wetted parts of the Hach Surface Scatter 7 sc turbidimeter are made with corrosion-resistant materials for extended life. The photo-detector and light source assemblies are protected from the effects of corrosive vapors and heated samples.

Wide turbidity measurement range from 0 to 9,999 NTU

Use the Surface Scatter 7 sc to establish and maintain optimum process control for reliable monitoring. Applications for this version include drinking water effluent, wastewater effluent, flocculation and sedimentation, industrial process water, and food processing waste containing fat, starch, or oil.

Wide turbidity measurement range from 0 to 9,999 NTU

The Surface Scatter 7sc turbidimeter can reliably measure turbidity from 0 to 9999 NTU in samples that vary from clear water to corrosive and high temperature paper mill and oil field samples.

Backed by a 2-year Warranty

Consistent with its long-standing reputation for quality and customer service, Hach Company warrants all Surface Scatter 7sc turbidimeters against defects in materials or workmanship for two years from the date of shipment.
 

Specifications

 

 
Technical Attributes
Accuracy±5% of reading or ±0.1 NTU (whichever is greater) from 0 - 2000 NTU
Accuracy 2±10% of reading from 2000 - 9999 NTU
AlarmRelays can be programmed through sc controller for warning, alarm, feeder, event, pulse width modulation, and frequency control.
Ambient Temperature0 - 50 °C
ChannelsInstrument Only
ControllerNo
Enclosure ConstructionPlastic instrument enclosure
Enclosure RatingNEMA 12
Installation CategoryCategory II
MountingWall mount
Number of ChannelsNone
Power Requirements (Voltage)None
Range0.1 - 9999 NTU
Relative Humidity5 - 95% non-condensing
Repeatability1.0% or ±0.04 NTU, whichever is greater
Resolution<1000 NTU: 0.01 NTU
1000 - 9999.9 NTU: 0.1 NTU
Response TimeInitial response in 45 seconds
Sample Flow RateSample flow required 1.0 - 2.0 L/min
Sample Input0.75 " NPT female thread
Sample Output1 " NPT female thread
Sample Temperature0 - 50°C (32 - 122°F)
Sample Temperature0 - 50 °C
Storage Conditions-20 - 80 °C
Warranty24 months
Weight15.9 kg
What's included?SS7®sc, Wall Mount Kit, Cylinder Brush, Light Source Alignment Template, 0.75 x 0.75 inch NPT Nipple, Calibration Cup, 0.75 inch NPT Ball Valve, SS7sc User Manual, and 500 mL 4000 NTU Formazin Standard.