Outstanding results in high temperatures: SURFACE SCATTER 7 sc
The Surface Scatter 7 sc High Range Turbidimeter with sc200 controller offers superior performance features. It measures a turbidity range from 0 to 9999 NTU. Includes calibration cylinder, 4000 NTU Formazin, installation accessories, and instruction manual. Note: only one SS7 sc sensor can attach to one sc200 controller, if a second different sc sensor is required for this configuration please contact the Hach Company - Process Technical Group for further details.
The Hach Surface Scatter 7 sc Turbidimeter (SS7) is uniquely designed so that the light source and photocell never come in contact with the sample. In fluids with high loads of suspended solids this makes sample cell cleaning and replacement unnecessary.
All wetted parts of the Hach Surface Scatter 7 sc turbidimeter are made with corrosion-resistant materials for extended life. The photo-detector and light source assemblies are protected from the effects of corrosive vapors and heated samples.
Use the Surface Scatter 7 sc to establish and maintain optimum process control for reliable monitoring. Applications for this version include drinking water effluent, wastewater effluent, flocculation and sedimentation, industrial process water, and food processing waste containing fat, starch, or oil.
The Surface Scatter 7sc turbidimeter can reliably measure turbidity from 0 to 9999 NTU in samples that vary from clear water to corrosive and high temperature paper mill and oil field samples.
Consistent with its long-standing reputation for quality and customer service, Hach Company warrants all Surface Scatter 7sc turbidimeters against defects in materials or workmanship for two years from the date of shipment.
Accuracy | ±5% of reading or ±0.1 NTU (whichever is greater) from 0 - 2000 NTU |
Accuracy 2 | ±10% of reading from 2000 - 9999 NTU |
Alarm | Relays can be programmed through sc controller for warning, alarm, feeder, event, pulse width modulation, and frequency control. |
Ambient Temperature | 0 - 50 °C |
Channels | Instrument Only |
Controller | No |
Enclosure Construction | Plastic instrument enclosure |
Enclosure Rating | NEMA 12 |
Installation Category | Category II |
Mounting | Wall mount |
Number of Channels | None |
Power Requirements (Voltage) | None |
Range | 0.1 - 9999 NTU |
Relative Humidity | 5 - 95% non-condensing |
Repeatability | 1.0% or ±0.04 NTU, whichever is greater |
Resolution | <1000 NTU: 0.01 NTU 1000 - 9999.9 NTU: 0.1 NTU |
Response Time | Initial response in 45 seconds |
Sample Flow Rate | Sample flow required 1.0 - 2.0 L/min |
Sample Input | 0.75 " NPT female thread |
Sample Output | 1 " NPT female thread |
Sample Temperature | 0 - 50°C (32 - 122°F) |
Sample Temperature | 0 - 50 °C |
Storage Conditions | -20 - 80 °C |
Warranty | 24 months |
Weight | 15.9 kg |
What's included? | SS7®sc, Wall Mount Kit, Cylinder Brush, Light Source Alignment Template, 0.75 x 0.75 inch NPT Nipple, Calibration Cup, 0.75 inch NPT Ball Valve, SS7sc User Manual, and 500 mL 4000 NTU Formazin Standard. |