Features:
1.Innovative door-free design for easy operation
2.Middle Detection and Ultrasound Detectors to double guarantee the correct IV set loading
3.Anti-bolus system to reduce significantly bolus after occlusion sudden release
4.Reminder alarm in case of forgetting to start the operation
5.Optional drip sensor and fluid heater
6.Wireless data transmission with infusion supervision system
7. Compact design with light weight
8. Embedded handle allows for easy transportation
9. Straightforward IV set installation
10. Intuitive programming procedure with quick guidance on screen
Specifications:Infusion set | Applicable to GB 8368-2005 disposals; diametersΦ3.5-4.5mm, thickness 0.8-1.2mm |
Flow rate range | 0.1~1500ml/h |
Accumulated volume | 0.1~9999ml |
VTBI | 0.1~999.9ml, increment is 0.1ml; 1000-9999ml, increment is 1ml |
Increment | 0.1~999.9ml/h, increment is 0.1ml/h; 1000~1500ml/h, increment is 1ml/h |
Accuracy | ≤±5% |
KVO rate | 0.5ml/h |
Bolus rate | 0.10-1500ml/h |
Purge rate | 600ml/h |
Occlusion | Low 150±125 mmHg (20kPa); Medium 525±125 mmHg (70kPa); High 900±180 mmHg (120 kPa) |
Air bubble detection | Single bubble detection: 50, 100, 250, 500, 800lμl Accumulated bubble detection: 0.1-4.0ml/h, adjustable |
Handle | Embedded handle |
Mounting type | Standard: Vertical Pole |
Language | 8, Chinese, English, Spanish, French, Russian, Turkish, Portuguese, German |
Classification | CF, Class I |
IP Grade | IP 34 |