Dialysis machine YJ-5000
Features: The blood purification equipment with full functions: CVVHDF, CPFA, DFPP, MARS, FPSA, HP … Full Range Of Blood Purification Equipments Citrate anti-coagulation and double input channel for substitution fluid flow |
Clinical advantages of citrate anticoagulation: * All treatment modes can adopt the Citrate anti-coagulation. * Seperate citrate infusion pump and weighing scale. * Control the citrate flow accurately. Automatic injection by proportion,achieve the safe topical citrate anticoagulation. |
Product parameters | |
Basic Information | |
Dimension | Height 1720 mm/Width 635 mm/Thickness 578 mm |
Weight | About 75Kg |
Electrical Parameters | |
Voltage | -110/220V |
Frequency | 50/60Hz |
Power | 500VA |
Backup Battery Capacity | 2×12V/4Ah |
Pressure Monitoring | |
Venous Pressure Monitoring(Vp) | -500mmHg~+700mmHg Accuracy:±5mmHg |
Arterial Pressure Monitoring(Ap) | -500mmHg~+700mmHg Accuracy:±5mmHg |
Transmembrane Pressure Monitoring(Tmp) | -500mmHg~+700mmHg Accuracy:±5mmHg |
Primary Membrane External Pressure Monitoring(Pm1) | Accuracy:±5mmHg -500mmHg~+700mmHg |
Secondary Membrane Pressure Monitoring(Pm2) | -500mmHg~+700mmHg Accuracy:±5mmHg |
Liquid Balance Control | |
ScaleNumber | 4 |
Rehydration Scale | Accuracy;±5gor ±0.1% Okg~30kg |
Waste Liquid Scale | Okg~30kg Accuracy:±5gor ±0.1% |
Auxiliary Rehydration Scalel | Okg~12kg Accuracy;±5gor±0.1% |
Flow Monitoring | |
Blood pump flow(BP) | 0mL/min,30mL/min~600mL/min Accuracy: set value±10% |
LP1 | 0mL/min,1ml/min~250ml/min Accuracy:±0.1mL/min or±5% ofsetvalue |
Waste Liquid Pump Flow(UFP) | 0mL/min,1mL/min~250mL/min Accuracy:±0.1mL/min or±5% ofsetvalue |
LP2 | 0mL/min,1mL/min ~250mL/min Accuracy:±0.1mL/min or±5% ofsetvalue |
Function Pump Flow(FP) | 0mL/min,1mL/min~250mL/min Accuracy:±0.1mL/minor±5% ofsetvalue |
Infusion pump flow | |
Flow Adjustable Range | 0mL/h~10mL/h |
Error Range | 0.2mL/hor±5% |
Rapid Injection Flow | 10mL/h~2000mL/h |
Syringe Specification | 10mL/20mL/30mL/50ml |
Dehydration Error | <±20mL/h |
Liquid Balance Control | 100ml |
Substitution fluid flow temperature control | |
Control Range | 33ºC~ 40ºC |
Accuracy | Control error ±1° C |
Safety Monitoring | |
Blood Leakage Monitoring | ≤ 0.35mL/min(HCT 32%) |
Air Monitoring | >0.02mL bubbles |
Note: The above parameters are YJ-5000 series, Specific parameters are subject to the operating manual. |